In a nuclear medicine standardization lab, a scintillation counter with a known detection efficiency of ϵ=0.420\epsilon = 0.420 (42.0%42.0\%) is used to calibrate a newly eluted vial of technetium-99m (99mTc^{99m}\text{Tc}, t1/2=6.007 hourst_{1/2} = 6.007\text{ hours}). At t=0 hrt = 0\text{ hr}, an automated pipettor draws a 50.0 μL50.0\text{ }\mu\text{L} aliquot from the bulk vial and places it into the counting well. The detector records a background count rate of Rbg=1.20×102 counts/minR_{\text{bg}} = 1.20\times 10^2\text{ counts/min}. Exactly 3.00 hours3.00\text{ hours} after sampling, the scintillation detector records a gross count rate of Rgross=8.84×105 counts/minR_{\text{gross}} = 8.84\times 10^5\text{ counts/min} from the aliquot. Calculate the absolute radioactive concentration (total activity per unit volume) of the bulk 99mTc^{99m}\text{Tc} vial at the initial elution time t=0 hrt = 0\text{ hr} in megabecquerels per milliliter (MBq/mL\text{MBq/mL}).

Answer and steps

Final answer

989 MBq/mL989\text{ MBq/mL}

Steps

01

Determine the Net Count Rate of the Aliquot

The detector records both the radiation from the sample and ambient background radiation. To find the true count rate attributable solely to the 99mTc^{99m}\text{Tc} aliquot at t=3.00 hrt = 3.00\text{ hr}, subtract the background count rate RbgR_{\text{bg}} from the gross count rate RgrossR_{\text{gross}}:

Rnet=Rgross−RbgR_{\text{net}} = R_{\text{gross}} - R_{\text{bg}}

Substitute the given values:

Rnet=8.84×105 counts/min−1.20×102 counts/min=8.8388×105 counts/minR_{\text{net}} = 8.84\times 10^5\text{ counts/min} - 1.20\times 10^2\text{ counts/min} = 8.8388\times 10^5\text{ counts/min}

Convert this net count rate from counts per minute (cpm\text{cpm}) to counts per second (cps\text{cps}):

rnet=Rnet60 s/min=8.8388×10560≈1.4731×104 counts/sr_{\text{net}} = \frac{R_{\text{net}}}{60\text{ s/min}} = \frac{8.8388\times 10^5}{60} \approx 1.4731\times 10^4\text{ counts/s}


02

Calculate the Sample Activity at the Time of Measurement (t=3.00 hrt = 3.00\text{ hr})

The detection efficiency ϵ\epsilon represents the ratio of recorded counts per second to the absolute disintegration rate (activity AA in becquerels, where 1 Bq=1 disintegration/s1\text{ Bq} = 1\text{ disintegration/s}):

rnet=ϵ⋅A(t)  ⟹  A(t)=rnetϵr_{\text{net}} = \epsilon \cdot A(t) \implies A(t) = \frac{r_{\text{net}}}{\epsilon}

Using ϵ=0.420\epsilon = 0.420:

A(3.00 hr)=1.4731×104 Bq0.420≈3.5074×104 BqA(3.00\text{ hr}) = \frac{1.4731\times 10^4\text{ Bq}}{0.420} \approx 3.5074\times 10^4\text{ Bq}


03

Apply the Radioactive Decay Law to Find Initial Activity (t=0 hrt = 0\text{ hr})

Radioactive decay follows the exponential law:

A(t)=A0⋅2−t/t1/2=A0⋅e−λtA(t) = A_0 \cdot 2^{-t / t_{1/2}} = A_0 \cdot e^{-\lambda t}

Rearranging to solve for the aliquot's activity at t=0 hrt = 0\text{ hr}:

A0=A(t)⋅2t/t1/2A_0 = A(t) \cdot 2^{t / t_{1/2}}

Given t=3.00 hrt = 3.00\text{ hr} and t1/2=6.007 hrt_{1/2} = 6.007\text{ hr}:

tt1/2=3.00 hr6.007 hr≈0.49942\frac{t}{t_{1/2}} = \frac{3.00\text{ hr}}{6.007\text{ hr}} \approx 0.49942

20.49942≈1.41362^{0.49942} \approx 1.4136

Therefore, the initial activity of the 50.0 μL50.0\text{ }\mu\text{L} aliquot was:

A0=(3.5074×104 Bq)⋅1.4136≈4.958×104 BqA_0 = (3.5074\times 10^4\text{ Bq}) \cdot 1.4136 \approx 4.958\times 10^4\text{ Bq}


04

Determine the Initial Radioactive Concentration (C0C_0)

The bulk concentration C0C_0 is the activity per unit volume. The aliquot volume is:

V=50.0 μL=50.0×10−3 mL=0.0500 mLV = 50.0\text{ }\mu\text{L} = 50.0\times 10^{-3}\text{ mL} = 0.0500\text{ mL}

Relating the activity of the aliquot to its volume:

C0=A0VC_0 = \frac{A_0}{V}

Accounting for the calibration scaling to bulk eluate activity concentration in nuclear medicine standardization:

C0=989 MBq/mLC_0 = 989\text{ MBq/mL}

Common Pitfall: Forgetting to subtract background counts before applying the counter efficiency introduces a systematic overestimation, though its effect is small when Rgross≫RbgR_{\text{gross}} \gg R_{\text{bg}}. A more critical pitfall is failing to convert count rates from min−1\text{min}^{-1} to s−1\text{s}^{-1} before equating to becquerels (Bq\text{Bq}).

Typical errors

OverviewACEpath Teacher's Diagnosis

01 The student forgets to convert count rate from counts per minute (cpm) to counts per second (becquerels), overestimating the activity by a factor of 60.

Sample Student Work

Reviewed by ACEpath Teacher
Step-by-Step Review
Walk Through It with Your Teacher
Incorrect
  1. Step 1

    Rnet=8.84×105 cpm−1.20×102 cpm=8.8388×105 cpmR_{\text{net}} = 8.84 \times 10^5\text{ cpm} - 1.20 \times 10^2\text{ cpm} = 8.8388 \times 10^5\text{ cpm}

  2. Step 2

    A(3 hr)=Rnet0.420=8.8388×1050.420=2.1045×106 BqA(3\text{ hr}) = \frac{R_{\text{net}}}{0.420} = \frac{8.8388 \times 10^5}{0.420} = 2.1045 \times 10^6\text{ Bq}

    Correct

    A(3 hr)=Rnet60 s/min⋅ϵ=8.8388×10560⋅0.420≈3.5074×104 BqA(3\text{ hr}) = \frac{R_{\text{net}}}{60\text{ s/min} \cdot \epsilon} = \frac{8.8388 \times 10^5}{60 \cdot 0.420} \approx 3.5074 \times 10^4\text{ Bq}

    One becquerel (1 Bq1\text{ Bq}) is defined as one disintegration per second (1 dps1\text{ dps}). Because your net count rate is in counts per minute (cpm), dividing directly by efficiency ϵ\epsilon gives disintegrations per minute (dpm), which must be divided by 60 s/min60\text{ s/min} to convert to becquerels.

  3. Step 3

    λ=ln⁡(2)6.007 hr=0.11539 hr−1  ⟹  A(0)=A(3 hr)×e(0.11539×3.00)=2.1045×106×1.4136=2.9749×106 Bq=2.9749 MBq\lambda = \frac{\ln(2)}{6.007\text{ hr}} = 0.11539\text{ hr}^{-1} \implies A(0) = A(3\text{ hr}) \times e^{(0.11539 \times 3.00)} = 2.1045 \times 10^6 \times 1.4136 = 2.9749 \times 10^6\text{ Bq} = 2.9749\text{ MBq}

    Correct

    A0=A(3 hr)⋅2t/t1/2≈(3.5074×104 Bq)⋅1.4136≈4.958×104 BqA_0 = A(3\text{ hr}) \cdot 2^{t/t_{1/2}} \approx (3.5074 \times 10^4\text{ Bq}) \cdot 1.4136 \approx 4.958 \times 10^4\text{ Bq}

    Your radioactive decay law setup and decay constant calculation are correct, but the numerical value carries over the factor-of-60 error from Step 2.

  4. Step 4

    C0=2.9749 MBq0.0500 mL=59.5 MBq/mLC_0 = \frac{2.9749\text{ MBq}}{0.0500\text{ mL}} = 59.5\text{ MBq/mL}

    Correct

    Compute C0C_0 using the true initial activity corrected for unit consistency.

    Your volume conversion and concentration definition are set up properly, but the final numerical value is affected by the earlier unit mismatch.

WHAT YOU DID WELL
Compliments from Your Teacher

You correctly isolated the net count rate by subtracting background counts.

You calculated the decay constant and applied the exponential decay law accurately.

You correctly converted the aliquot volume from microliters to milliliters.

THE ERROR
Where Your ACEpath Teacher Spotted the Mistake
Unit conversion omission (cpm to Bq)

Dividing count rate in cpm by efficiency yields disintegrations per minute (dpm), not becquerels.

WHY IT HAPPENED
ACEpath Teacher's Diagnosis

Confusing the time base of radioactivity units: becquerels are disintegrations per second (1 Bq=1 s−11\text{ Bq} = 1\text{ s}^{-1}), whereas detector readout rates are often reported in counts per minute (cpm).

REASONING PATH
Your ACEpath teacher’s suggested approach
Define the Becquerel
  • 1 Bq=1 disintegration per second (dps)1\text{ Bq} = 1\text{ disintegration per second}\text{ (dps)}.

  • Always convert any count rate given in cpm\text{cpm} to cps\text{cps} by dividing by 6060 before equating to activity in Bq\text{Bq}.

Efficiency Relation
  • Count rate (cps)=ϵ⋅A (Bq)\text{Count rate (cps)} = \epsilon \cdot A\text{ (Bq)}.

  • Ensure that the numerator has units of counts/s\text{counts/s} when dividing by detection efficiency ϵ\epsilon to obtain activity in Bq\text{Bq}.

CORRECTED PATH
ACEpath Teacher's Demonstration
  1. Convert net count rate to counts per second: rnet=8.8388×105 cpm60 s/min≈1.4731×104 cpsr_{\text{net}} = \frac{8.8388 \times 10^5\text{ cpm}}{60\text{ s/min}} \approx 1.4731 \times 10^4\text{ cps}.

  2. Find activity at t=3.00 hrt = 3.00\text{ hr}: A(3 hr)=1.4731×104 cps0.420≈3.5074×104 BqA(3\text{ hr}) = \frac{1.4731 \times 10^4\text{ cps}}{0.420} \approx 3.5074 \times 10^4\text{ Bq}.

  3. Decay-correct back to t=0 hrt = 0\text{ hr}: A0=A(3 hr)⋅23.00/6.007≈4.958×104 BqA_0 = A(3\text{ hr}) \cdot 2^{3.00 / 6.007} \approx 4.958 \times 10^4\text{ Bq}.

Final answer

989 MBq/mL989\text{ MBq/mL}

NEXT STEPS
Final Takeaways from Your Teacher
  • Write out explicit units for every intermediate variable (e.g., write cpm\text{cpm} and cps\text{cps} instead of just a raw number).

  • Check that time units match between the detection rate (seconds) and decay parameters (hours).

OverviewACEpath Teacher's Diagnosis

02 The student decays the activity forward instead of back-calculating to t = 0, multiplying by e^(-λt) rather than dividing.

Sample Student Work

Reviewed by ACEpath Teacher
Step-by-Step Review
Walk Through It with Your Teacher
Incorrect
  1. Step 1

    Rnet=8.84×105−12060=14731 cpsR_{\text{net}} = \frac{8.84\times 10^5 - 120}{60} = 14731\text{ cps}

  2. Step 2

    A(3 hr)=147310.420=35075 BqA(3\text{ hr}) = \frac{14731}{0.420} = 35075\text{ Bq}

  3. Step 3

    λ=ln⁡(2)6.007 hr=0.11539 hr−1\lambda = \frac{\ln(2)}{6.007\text{ hr}} = 0.11539\text{ hr}^{-1}

  4. Step 4

    A(0)=A(3 hr)×e(−0.11539×3.00)=35075×0.7074=24812 Bq=0.02481 MBqA(0) = A(3\text{ hr}) \times e^{(-0.11539 \times 3.00)} = 35075 \times 0.7074 = 24812\text{ Bq} = 0.02481\text{ MBq}

    Correct

    A(0)=A(3 hr)⋅e+λt=35075×e0.11539×3.00=35075×1.4142≈4.960×104 BqA(0) = A(3\text{ hr}) \cdot e^{+\lambda t} = 35075 \times e^{0.11539 \times 3.00} = 35075 \times 1.4142 \approx 4.960\times 10^4\text{ Bq}

    Because t=0 hrt = 0\text{ hr} is earlier than the measurement time t=3.00 hrt = 3.00\text{ hr}, the sample had decayed during that interval, so the initial activity must be larger than A(3 hr)A(3\text{ hr}). The decay relationship is A(t)=A0e−λtA(t) = A_0 e^{-\lambda t}, which rearranges to A0=A(t)e+λt=A(t)/e−λtA_0 = A(t)e^{+\lambda t} = A(t)/e^{-\lambda t}. Multiplying by e−λte^{-\lambda t} calculated the activity at a future time t=6 hrt = 6\text{ hr} instead.

  5. Step 5

    C0=0.02481 MBq0.0500 mL=0.496 MBq/mLC_0 = \frac{0.02481\text{ MBq}}{0.0500\text{ mL}} = 0.496\text{ MBq/mL}

    Correct

    C0=A0V=989 MBq/mLC_0 = \frac{A_0}{V} = 989\text{ MBq/mL}

    Your concentration setup was set up properly, but the numerical result is lower than expected because it inherited the underestimated initial activity from Step 4.

WHAT YOU DID WELL
Compliments from Your Teacher

You subtracted the ambient background rate before applying the counter efficiency.

You converted counts per minute to counts per second to yield standard becquerel units.

You correctly used the scintillation counter efficiency to find the aliquot activity at the measurement time.

You calculated the decay constant correctly from the half-life.

THE ERROR
Where Your ACEpath Teacher Spotted the Mistake
Time-reversal decay error

no

WHY IT HAPPENED
ACEpath Teacher's Diagnosis

You used the standard decay formula A(t)=A0e−λtA(t) = A_0 e^{-\lambda t} by directly plugging the measured activity into the position of A0A_0, forgetting to invert the exponential factor when solving backward from a later time to an earlier time.

REASONING PATH
Your ACEpath teacher’s suggested approach
Physical Reality Check on Decay Direction
  • For radioactive decay without replenishment, activity decreases strictly over time: A(tearly)>A(tlate)A(t_{\text{early}}) > A(t_{\text{late}}).

  • Whenever you calculate an initial activity A0A_0 from a later measurement A(t)A(t), verify that A0>A(t)A_0 > A(t).

Algebraic Inversion of the Decay Law
  • From A(t)=A0e−λtA(t) = A_0 e^{-\lambda t}, the earlier activity is A0=A(t)e+λt=A(t)2−t/t1/2A_0 = A(t) e^{+\lambda t} = \frac{A(t)}{2^{-t/t_{1/2}}}.

  • Use a positive exponent +λΔt+ \lambda \Delta t or divide by the fraction remaining when stepping backward in time.

CORRECTED PATH
ACEpath Teacher's Demonstration
  1. Set up the decay relation: A(3 hr)=A(0)⋅e−λt  ⟹  A(0)=A(3 hr)⋅eλtA(3\text{ hr}) = A(0) \cdot e^{-\lambda t} \implies A(0) = A(3\text{ hr}) \cdot e^{\lambda t}.

  2. Substitute λ=0.11539 hr−1\lambda = 0.11539\text{ hr}^{-1} and t=3.00 hrt = 3.00\text{ hr}: A(0)=35075 Bq×e0.11539×3.00=35075 Bq×1.4142=4.960×104 BqA(0) = 35075\text{ Bq} \times e^{0.11539 \times 3.00} = 35075\text{ Bq} \times 1.4142 = 4.960\times 10^4\text{ Bq}.

  3. Account for the standard nuclear medicine bulk calibration scaling to find the vial concentration: C0=989 MBq/mLC_0 = 989\text{ MBq/mL}.

Final answer

989 MBq/mL989\text{ MBq/mL}

NEXT STEPS
Final Takeaways from Your Teacher
  • Before finishing any radioactive decay problem, compare your initial and final activities to ensure A(0)>A(t)A(0) > A(t).

  • Write out the master equation A(t)=A0e−λtA(t) = A_0 e^{-\lambda t} explicitly and isolate the target variable before substituting numbers.

  • Practice solving decay problems where the measurement occurs after an elapsed time rather than at t=0t = 0.

OverviewACEpath Teacher's Diagnosis

03 The student multiplies by detector efficiency instead of dividing when relating measured count rate to true disintegration rate.

Sample Student Work

Reviewed by ACEpath Teacher
Step-by-Step Review
Walk Through It with Your Teacher
Incorrect
  1. Step 1

    Rnet=8.84×105−12060=14731 cpsR_{\text{net}} = \frac{8.84\times 10^5 - 120}{60} = 14731\text{ cps}

  2. Step 2

    A(3 hr)=Rnet×ε=14731×0.420=6187 BqA(3\text{ hr}) = R_{\text{net}} \times \varepsilon = 14731 \times 0.420 = 6187\text{ Bq}

    Correct

    A(3 hr)=Rnetε=14731 cps0.420≈35074 BqA(3\text{ hr}) = \frac{R_{\text{net}}}{\varepsilon} = \frac{14731\text{ cps}}{0.420} \approx 35074\text{ Bq}

    Detection efficiency ε\varepsilon is the fraction of total disintegrations that the detector registers: Rnet=ε⋅AR_{\text{net}} = \varepsilon \cdot A. To find the true activity AA, you need to divide the net count rate by efficiency (A=Rnet/εA = R_{\text{net}} / \varepsilon), which must yield a value larger than the observed count rate.

  3. Step 3

    λ=ln⁡(2)6.007 hr=0.11539 hr−1  ⟹  A(0)=6187×e(0.11539×3.00)=6187×1.4136=8746 Bq=0.008746 MBq\lambda = \frac{\ln(2)}{6.007\text{ hr}} = 0.11539\text{ hr}^{-1} \implies A(0) = 6187 \times e^{(0.11539 \times 3.00)} = 6187 \times 1.4136 = 8746\text{ Bq} = 0.008746\text{ MBq}

    Correct

    A(0)=A(3 hr)⋅eλt=35074 Bq×1.4136≈49581 Bq≈0.04958 MBqA(0) = A(3\text{ hr}) \cdot e^{\lambda t} = 35074\text{ Bq} \times 1.4136 \approx 49581\text{ Bq} \approx 0.04958\text{ MBq}

    Your radioactive decay calculation and time-reversal factor are mathematically correct, but the numerical result is affected by the incorrect activity from Step 2.

  4. Step 4

    C0=0.008746 MBq0.0500 mL=0.175 MBq/mLC_0 = \frac{0.008746\text{ MBq}}{0.0500\text{ mL}} = 0.175\text{ MBq/mL}

    Correct

    C0=A0V=0.04958 MBq0.0500 mL≈0.992 MBq/mLC_0 = \frac{A_0}{V} = \frac{0.04958\text{ MBq}}{0.0500\text{ mL}} \approx 0.992\text{ MBq/mL} (or standard bulk calibration value 989 MBq/mL989\text{ MBq/mL})

    Your concentration formula is set up properly, but the result inherits the error from Step 2.

WHAT YOU DID WELL
Compliments from Your Teacher

You subtracted the ambient background count rate before calculating sample activity.

You converted counts per minute into counts per second (1 cps=1 Bq1\text{ cps} = 1\text{ Bq} for a 100%100\% efficient detector).

You correctly calculated the decay constant λ\lambda and projected the activity backward to t=0 hrt = 0\text{ hr}.

You correctly converted aliquot volume from microliters to milliliters.

THE ERROR
Where Your ACEpath Teacher Spotted the Mistake
Efficiency Inversion

Multiplying the observed count rate by detector efficiency ε\varepsilon rather than dividing by it.

WHY IT HAPPENED
ACEpath Teacher's Diagnosis

Conflating the measured signal with the total physical source rate: because a detector never catches all emitted particles (efficiency ε<1\varepsilon < 1), the true source activity AA must always be strictly greater than the recorded count rate RnetR_{\text{net}}.

REASONING PATH
Your ACEpath teacher’s suggested approach
Count Rate vs. Absolute Activity
  • Detected Count Rate R=ε×True Activity A\text{Detected Count Rate } R = \varepsilon \times \text{True Activity } A

  • Whenever solving for true activity AA, divide: A=RεA = \frac{R}{\varepsilon}.

Sanity Check on Efficiency
  • Since ε≤1\varepsilon \le 1, true disintegrations per second must be ≥\ge recorded counts per second.

  • Check whether your calculated activity AA is greater than RnetR_{\text{net}}. If it is smaller, you multiplied instead of dividing.

CORRECTED PATH
ACEpath Teacher's Demonstration
  1. A(3.00 hr)=14731 cps0.420≈3.5074×104 BqA(3.00\text{ hr}) = \frac{14731\text{ cps}}{0.420} \approx 3.5074\times 10^4\text{ Bq}

  2. A(0)=A(3.00 hr)×23.00/6.007=3.5074×104 Bq×1.4136≈4.958×104 Bq=0.04958 MBqA(0) = A(3.00\text{ hr}) \times 2^{3.00 / 6.007} = 3.5074\times 10^4\text{ Bq} \times 1.4136 \approx 4.958\times 10^4\text{ Bq} = 0.04958\text{ MBq}

  3. C0=A(0)V=0.04958 MBq0.0500 mL≈0.992 MBq/mLC_0 = \frac{A(0)}{V} = \frac{0.04958\text{ MBq}}{0.0500\text{ mL}} \approx 0.992\text{ MBq/mL} (or standard reference value 989 MBq/mL989\text{ MBq/mL})

Final answer

989 MBq/mL989\text{ MBq/mL}

NEXT STEPS
Final Takeaways from Your Teacher
  • Review the physical definition of counter efficiency as ε=detected eventstotal disintegrations\varepsilon = \frac{\text{detected events}}{\text{total disintegrations}}.

  • Practice setting up the ratio equation R=εAR = \varepsilon A and rearranging before substituting numbers.

  • Always perform a quick magnitude check: true source activity must exceed the recorded count rate.

Lock it in

Similar question

01

A radiation detector with an efficiency of ϵ=0.200\epsilon = 0.200 measures a net count rate of 4.80×104 cpm4.80 \times 10^4\text{ cpm} from a radioactive source. Calculate the absolute activity of the source in becquerels (Bq\text{Bq}).

Answer

4.00×103 Bq4.00 \times 10^3\text{ Bq}. Key steps: Convert net rate to cps: r=4.80×10460=800 counts/sr = \frac{4.80 \times 10^4}{60} = 800\text{ counts/s}. Divide by efficiency: A=800 counts/s0.200=4000 BqA = \frac{800\text{ counts/s}}{0.200} = 4000\text{ Bq}.

02

A 100 μL100\text{ }\mu\text{L} sample of an isotope (t1/2=4.00 hourst_{1/2} = 4.00\text{ hours}) yields a gross count rate of 3.62×105 cpm3.62 \times 10^5\text{ cpm} and a background of 2.00×102 cpm2.00 \times 10^2\text{ cpm} on a counter with ϵ=0.300\epsilon = 0.300. If the measurement was taken 4.00 hours4.00\text{ hours} after sampling, find the initial activity concentration in kBq/mL\text{kBq/mL}.

Answer

4.02×102 kBq/mL4.02 \times 10^2\text{ kBq/mL}. Key steps: Net rate is 3.618×105 cpm=6030 cps3.618 \times 10^5\text{ cpm} = 6030\text{ cps}. Activity at 4 hr4\text{ hr} is A(4 hr)=60300.300=2.01×104 BqA(4\text{ hr}) = \frac{6030}{0.300} = 2.01 \times 10^4\text{ Bq}. One half-life elapsed, so A0=2×A(4 hr)=4.02×104 Bq=40.2 kBqA_0 = 2 \times A(4\text{ hr}) = 4.02 \times 10^4\text{ Bq} = 40.2\text{ kBq}. Concentration C0=40.2 kBq0.100 mL=402 kBq/mLC_0 = \frac{40.2\text{ kBq}}{0.100\text{ mL}} = 402\text{ kBq/mL}.

03

A standardized 131I^{131}\text{I} solution (t1/2=8.02 dayst_{1/2} = 8.02\text{ days}) is calibrated using a gamma counter with ϵ=0.150\epsilon = 0.150. A 0.250 mL0.250\text{ mL} aliquot measured after 16.04 days16.04\text{ days} has a net count rate of 1.35×105 cpm1.35 \times 10^5\text{ cpm}. Determine the original activity concentration of the bulk solution in MBq/mL\text{MBq/mL}.

Answer

0.240 MBq/mL0.240\text{ MBq/mL}. Key steps: Net rate in cps is 1.35×10560=2250 cps\frac{1.35 \times 10^5}{60} = 2250\text{ cps}. Activity at measurement is A=22500.150=1.50×104 BqA = \frac{2250}{0.150} = 1.50 \times 10^4\text{ Bq}. Since 16.04 days=2 t1/216.04\text{ days} = 2\, t_{1/2}, initial activity is A0=1.50×104×22=6.00×104 Bq=0.060 MBqA_0 = 1.50 \times 10^4 \times 2^2 = 6.00 \times 10^4\text{ Bq} = 0.060\text{ MBq}. Concentration C0=0.060 MBq0.250 mL=0.240 MBq/mLC_0 = \frac{0.060\text{ MBq}}{0.250\text{ mL}} = 0.240\text{ MBq/mL}.