An engineering student calibrates an automated capillary viscometer to measure the dynamic viscosity μ\mu of an engine lubricant. Poiseuille's law governs the laminar flow through the horizontal capillary tube: ΔP=8μLQπR4\Delta P = \frac{8 \mu L Q}{\pi R^4}, where ΔP\Delta P is pressure drop across the tube length LL, RR is internal radius, and QQ is volumetric flow rate. In the lab trial, a tube of length L=15.0 cmL = 15.0\text{ cm} and internal diameter d=2.00 mmd = 2.00\text{ mm} carries an oil flow of Q=1.80 mL/minQ = 1.80\text{ mL/min} under an applied gauge pressure differential of ΔP=4.50 kPa\Delta P = 4.50\text{ kPa}. Using dimensional analysis and consistent SI base units, isolate μ\mu and calculate the dynamic viscosity of the oil in both pascal-seconds (Pa⋅s\text{Pa}\cdot\text{s}) and centipoise (cP\text{cP}), where 1 cP=10−3 Pa⋅s=10−3 kg/(m⋅s)1\text{ cP} = 10^{-3}\text{ Pa}\cdot\text{s} = 10^{-3}\text{ kg/(m}\cdot\text{s)}. Express your answer to three significant figures.

Answer and steps

Final answer

μ=0.393 Pa⋅s=393 cP\mu = 0.393\text{ Pa}\cdot\text{s} = 393\text{ cP}

Steps

01

Isolate the Dynamic Viscosity μ\mu

Starting from Poiseuille's law: ΔP=8μLQπR4\Delta P = \frac{8 \mu L Q}{\pi R^4}

Multiply both sides by πR4\pi R^4 and divide by 8LQ8 L Q to solve for dynamic viscosity μ\mu: μ=πR4ΔP8LQ\mu = \frac{\pi R^4 \Delta P}{8 L Q}


02

Verify Units via Dimensional Analysis

Substitute the SI dimensions into the rearranged equation:

  • [ΔP]=Pa=Nm2=kgm⋅s2[\Delta P] = \text{Pa} = \frac{\text{N}}{\text{m}^2} = \frac{\text{kg}}{\text{m}\cdot\text{s}^2}
  • [R4]=m4[R^4] = \text{m}^4
  • [L]=m[L] = \text{m}
  • [Q]=m3s[Q] = \frac{\text{m}^3}{\text{s}}

[μ]=m4⋅Pam⋅(m3s)=m4⋅Pam4s=Pa⋅s[\mu] = \frac{\text{m}^4 \cdot \text{Pa}}{\text{m} \cdot \left(\frac{\text{m}^3}{\text{s}}\right)} = \frac{\text{m}^4 \cdot \text{Pa}}{\frac{\text{m}^4}{\text{s}}} = \text{Pa}\cdot\text{s}

Expressed in SI base units: Pa⋅s=(kgm⋅s2)⋅s=kgm⋅s\text{Pa}\cdot\text{s} = \left(\frac{\text{kg}}{\text{m}\cdot\text{s}^2}\right)\cdot\text{s} = \frac{\text{kg}}{\text{m}\cdot\text{s}}

This confirms that the formula yields the correct dimensional unit for dynamic viscosity.


03

Convert All Given Values to SI Base Units

  • Radius (RR): The problem provides the internal diameter d=2.00 mmd = 2.00\text{ mm}.
    R=d2=2.00 mm2=1.00 mm=1.00×10−3 mR = \frac{d}{2} = \frac{2.00\text{ mm}}{2} = 1.00\text{ mm} = 1.00 \times 10^{-3}\text{ m} (Common pitfall: plugging in the diameter instead of the radius will lead to an error by a factor of 24=162^4 = 16.)
  • Length (LL): L=15.0 cm=15.0×10−2 m=0.150 mL = 15.0\text{ cm} = 15.0 \times 10^{-2}\text{ m} = 0.150\text{ m}
  • Pressure drop (ΔP\Delta P): ΔP=4.50 kPa=4.50×103 Pa\Delta P = 4.50\text{ kPa} = 4.50 \times 10^3\text{ Pa}
  • Volumetric flow rate (QQ):
    Recall that 1 mL=1 cm3=10−6 m31\text{ mL} = 1\text{ cm}^3 = 10^{-6}\text{ m}^3 and 1 min=60 s1\text{ min} = 60\text{ s}: Q=1.80 mL1 min=1.80×10−6 m360 s=3.00×10−8 m3/sQ = \frac{1.80\text{ mL}}{1\text{ min}} = \frac{1.80 \times 10^{-6}\text{ m}^3}{60\text{ s}} = 3.00 \times 10^{-8}\text{ m}^3/\text{s}

04

Calculate the Numerical Value

Substitute the converted values into the formula: μ=π⋅(1.00×10−3 m)4⋅(4.50×103 Pa)8⋅(0.150 m)⋅(3.00×10−8 m3/s)\mu = \frac{\pi \cdot (1.00 \times 10^{-3}\text{ m})^4 \cdot (4.50 \times 10^3\text{ Pa})}{8 \cdot (0.150\text{ m}) \cdot (3.00 \times 10^{-8}\text{ m}^3/\text{s})}

Calculate the numerator: π⋅(1.00×10−12 m4)⋅(4.50×103 Pa)=4.50π×10−9 Pa⋅m4\pi \cdot (1.00 \times 10^{-12}\text{ m}^4) \cdot (4.50 \times 10^3\text{ Pa}) = 4.50\pi \times 10^{-9}\text{ Pa}\cdot\text{m}^4

Calculate the denominator: 8⋅0.150 m⋅(3.00×10−8 m3/s)=1.20⋅3.00×10−8=3.60×10−8 m4/s8 \cdot 0.150\text{ m} \cdot (3.00 \times 10^{-8}\text{ m}^3/\text{s}) = 1.20 \cdot 3.00 \times 10^{-8} = 3.60 \times 10^{-8}\text{ m}^4/\text{s}

Divide numerator by denominator: μ=4.50π×10−93.60×10−8=4.50π36.0=π8 Pa⋅s≈0.392699 Pa⋅s\mu = \frac{4.50\pi \times 10^{-9}}{3.60 \times 10^{-8}} = \frac{4.50\pi}{36.0} = \frac{\pi}{8}\text{ Pa}\cdot\text{s} \approx 0.392699\text{ Pa}\cdot\text{s}

To three significant figures: μ=0.393 Pa⋅s\mu = 0.393\text{ Pa}\cdot\text{s}


05

Convert to Centipoise (cP\text{cP})

Using the relation 1 cP=10−3 Pa⋅s1\text{ cP} = 10^{-3}\text{ Pa}\cdot\text{s} (or 1 Pa⋅s=1000 cP1\text{ Pa}\cdot\text{s} = 1000\text{ cP}): μ=0.392699 Pa⋅s×1 cP10−3 Pa⋅s=392.699 cP≈393 cP\mu = 0.392699\text{ Pa}\cdot\text{s} \times \frac{1\text{ cP}}{10^{-3}\text{ Pa}\cdot\text{s}} = 392.699\text{ cP} \approx 393\text{ cP}

Typical errors

OverviewACEpath Teacher's Diagnosis

01 The student substitutes the diameter directly into Poiseuille's formula instead of dividing by 2 to find the radius, leading to a viscosity that is 16 times too large.

Sample Student Work

Reviewed by ACEpath Teacher
Step-by-Step Review
Walk Through It with Your Teacher
Incorrect
  1. Step 1

    μ=π⋅R4⋅ΔP8⋅L⋅Q\mu = \frac{\pi \cdot R^4 \cdot \Delta P}{8 \cdot L \cdot Q}

  2. Step 2

    L=0.150 m  ⟹  R=2.00 mm=2.00×10−3 m  ⟹  ΔP=4500 Pa  ⟹  Q=1.80×10−6/60=3.00×10−8 m3/sL = 0.150\text{ m} \implies R = 2.00\text{ mm} = 2.00 \times 10^{-3}\text{ m} \implies \Delta P = 4500\text{ Pa} \implies Q = 1.80 \times 10^{-6}/60 = 3.00 \times 10^{-8}\text{ m}^3/\text{s}

    Correct

    R=d2=2.00 mm2=1.00 mm=1.00×10−3 mR = \frac{d}{2} = \frac{2.00\text{ mm}}{2} = 1.00\text{ mm} = 1.00 \times 10^{-3}\text{ m}

    The problem provides the internal diameter d=2.00 mmd = 2.00\text{ mm}, not the radius. You must divide the diameter by 22 to obtain the radius R=1.00×10−3 mR = 1.00 \times 10^{-3}\text{ m}.

  3. Step 3

    μ=π⋅(2.00×10−3)4⋅45008⋅0.150⋅3.00×10−8=6.28 Pa⋅s\mu = \frac{\pi \cdot (2.00 \times 10^{-3})^4 \cdot 4500}{8 \cdot 0.150 \cdot 3.00 \times 10^{-8}} = 6.28\text{ Pa}\cdot\text{s}

    Correct

    μ=π⋅(1.00×10−3 m)4⋅(4.50×103 Pa)8⋅(0.150 m)⋅(3.00×10−8 m3/s)≈0.393 Pa⋅s\mu = \frac{\pi \cdot (1.00 \times 10^{-3}\text{ m})^4 \cdot (4.50 \times 10^3\text{ Pa})}{8 \cdot (0.150\text{ m}) \cdot (3.00 \times 10^{-8}\text{ m}^3/\text{s})} \approx 0.393\text{ Pa}\cdot\text{s}

    Your arithmetic was executed correctly based on your substituted values, but substituting R=2.00×10−3 mR = 2.00 \times 10^{-3}\text{ m} rather than 1.00×10−3 m1.00 \times 10^{-3}\text{ m} scaled the result by 24=162^4 = 16.

  4. Step 4

    μ=6280 cP\mu = 6280\text{ cP}

    Correct

    μ=0.393 Pa⋅s×1000 cP1 Pa⋅s=393 cP\mu = 0.393\text{ Pa}\cdot\text{s} \times \frac{1000\text{ cP}}{1\text{ Pa}\cdot\text{s}} = 393\text{ cP}

    You applied the conversion 1 Pa⋅s=1000 cP1\text{ Pa}\cdot\text{s} = 1000\text{ cP} correctly to your calculated value, but the result carries the earlier radius error.

WHAT YOU DID WELL
Compliments from Your Teacher

You correctly manipulated Poiseuille's law to isolate the dynamic viscosity μ\mu.

You converted length from centimeters to meters and pressure from kilopascals to pascals without error.

You correctly converted the flow rate from mL/min\text{mL/min} to m3/s\text{m}^3/\text{s}.

You used the correct conversion factor between Pa⋅s\text{Pa}\cdot\text{s} and cP\text{cP}.

THE ERROR
Where Your ACEpath Teacher Spotted the Mistake
Radius vs. Diameter Confusion

You labeled the given diameter d=2.00 mmd = 2.00\text{ mm} as the radius RR, skipping the division by 22.

WHY IT HAPPENED
ACEpath Teacher's Diagnosis

Overlooking whether a geometric cylindrical dimension is specified as diameter or radius before substituting into power-law formulas.

REASONING PATH
Your ACEpath teacher’s suggested approach
Check Diameter vs. Radius
  • Geometric flow formulas such as Poiseuille's law use radius R=d2R = \frac{d}{2}, whereas lab specs often report internal diameter dd.

  • Always circle whether a problem gives dd or RR and immediately write R=d/2R = d/2 before writing down conversions.

Evaluate Sensitivity to Powers
  • Because RR appears as R4R^4, using dd instead of RR introduces a relative error factor of 24=162^4 = 16.

  • Whenever a variable has an exponent greater than 11, double-check the base value before evaluating.

CORRECTED PATH
ACEpath Teacher's Demonstration
  1. Calculate the internal radius: R=d2=2.00×10−3 m2=1.00×10−3 mR = \frac{d}{2} = \frac{2.00 \times 10^{-3}\text{ m}}{2} = 1.00 \times 10^{-3}\text{ m}.

  2. Substitute the corrected radius into the formula: μ=π⋅(1.00×10−3 m)4⋅4500 Pa8⋅0.150 m⋅3.00×10−8 m3/s=4.50π×10−93.60×10−8=π8≈0.3927 Pa⋅s\mu = \frac{\pi \cdot (1.00 \times 10^{-3}\text{ m})^4 \cdot 4500\text{ Pa}}{8 \cdot 0.150\text{ m} \cdot 3.00 \times 10^{-8}\text{ m}^3/\text{s}} = \frac{4.50\pi \times 10^{-9}}{3.60 \times 10^{-8}} = \frac{\pi}{8} \approx 0.3927\text{ Pa}\cdot\text{s}.

  3. Round to three significant figures: μ=0.393 Pa⋅s\mu = 0.393\text{ Pa}\cdot\text{s}.

  4. Convert to centipoise: μ=0.3927 Pa⋅s×1000 cP/(Pa⋅s)≈393 cP\mu = 0.3927\text{ Pa}\cdot\text{s} \times 1000\text{ cP}/(\text{Pa}\cdot\text{s}) \approx 393\text{ cP}.

Final answer

μ=0.393 Pa⋅s=393 cP\mu = 0.393\text{ Pa}\cdot\text{s} = 393\text{ cP}

NEXT STEPS
Final Takeaways from Your Teacher
  • Review the distinction between diameter and radius in fluid dynamics and cross-sectional geometry formulas.

  • Practice checking high-exponent variables (r2r^2, r3r^3, r4r^4) to ensure the radius rather than diameter is used.

OverviewACEpath Teacher's Diagnosis

02 The student forgets to convert the volumetric flow rate from per minute to per second, causing the calculated viscosity to be 60 times too small.

Sample Student Work

Reviewed by ACEpath Teacher
Step-by-Step Review
Walk Through It with Your Teacher
Incorrect
  1. Step 1

    μ=π⋅R4⋅ΔP8⋅L⋅Q\mu = \frac{\pi \cdot R^4 \cdot \Delta P}{8 \cdot L \cdot Q}

  2. Step 2

    L=0.150 m  ⟹  R=1.00×10−3 m  ⟹  ΔP=4500 Pa  ⟹  Q=1.80×10−6 m3/sL = 0.150\text{ m} \implies R = 1.00 \times 10^{-3}\text{ m} \implies \Delta P = 4500\text{ Pa} \implies Q = 1.80 \times 10^{-6}\text{ m}^3/\text{s}

    Correct

    L=0.150 mL = 0.150\text{ m}, R=1.00×10−3 mR = 1.00 \times 10^{-3}\text{ m}, ΔP=4.50×103 Pa\Delta P = 4.50 \times 10^3\text{ Pa}, and Q=1.80×10−6 m360 s=3.00×10−8 m3/sQ = \frac{1.80 \times 10^{-6}\text{ m}^3}{60\text{ s}} = 3.00 \times 10^{-8}\text{ m}^3/\text{s}

    You converted LL, RR, and ΔP\Delta P accurately, and converted milliliters to cubic meters (1 mL=10−6 m31\text{ mL} = 10^{-6}\text{ m}^3). However, the flow rate was given per minute (/min/\text{min}), so you needed to divide by 60 s60\text{ s} to get the rate per second.

  3. Step 3

    μ=π⋅(1.00×10−3)4⋅45008⋅0.150⋅1.80×10−6=0.00654 Pa⋅s\mu = \frac{\pi \cdot (1.00 \times 10^{-3})^4 \cdot 4500}{8 \cdot 0.150 \cdot 1.80 \times 10^{-6}} = 0.00654\text{ Pa}\cdot\text{s}

    Correct

    μ=π⋅(1.00×10−3 m)4⋅(4500 Pa)8⋅(0.150 m)⋅(3.00×10−8 m3/s)≈0.393 Pa⋅s\mu = \frac{\pi \cdot (1.00 \times 10^{-3}\text{ m})^4 \cdot (4500\text{ Pa})}{8 \cdot (0.150\text{ m}) \cdot (3.00 \times 10^{-8}\text{ m}^3/\text{s})} \approx 0.393\text{ Pa}\cdot\text{s}

    Your arithmetic evaluation was correct for the substituted numbers, but the result is 60 times too low due to the flow rate error in Step 2.

  4. Step 4

    μ=6.54 cP\mu = 6.54\text{ cP}

    Correct

    μ=0.393 Pa⋅s×1 cP10−3 Pa⋅s=393 cP\mu = 0.393\text{ Pa}\cdot\text{s} \times \frac{1\text{ cP}}{10^{-3}\text{ Pa}\cdot\text{s}} = 393\text{ cP}

    You correctly multiplied by 1000 to convert from Pa⋅s\text{Pa}\cdot\text{s} to cP\text{cP}, but the numeric value remains affected by the earlier flow rate error.

WHAT YOU DID WELL
Compliments from Your Teacher

You correctly solved Poiseuille's formula algebraically for μ\mu.

You correctly halved the internal diameter to obtain the radius before converting to meters.

You correctly converted the pressure differential from kilopascals to pascals.

You used the correct conversion factor between Pa⋅s\text{Pa}\cdot\text{s} and cP\text{cP}.

THE ERROR
Where Your ACEpath Teacher Spotted the Mistake
Compound Unit Time Conversion Omission

You converted volume from milliliters to cubic meters correctly, but carried the per-minute quantity directly into seconds without dividing by 60.

WHY IT HAPPENED
ACEpath Teacher's Diagnosis

Incomplete compound unit conversion: when converting a rate unit like mL/min\text{mL/min} into the coherent SI base rate m3/s\text{m}^3/\text{s}, both the numerator (volume) and denominator (time) must be explicitly converted using dimensional factor-label steps.

REASONING PATH
Your ACEpath teacher’s suggested approach
Compound Rate Decomposition
  • For any rate quantity AB\frac{A}{B}, convert AA and BB independently using conversion factors: unitAunitB=(SIAunitA)⋅(1unitB→SIB)\frac{\text{unit}_A}{\text{unit}_B} = \left(\frac{\text{SI}_A}{\text{unit}_A}\right) \cdot \left(\frac{1}{\text{unit}_B \to \text{SI}_B}\right).

  • Write out Q=1.80mLmin×10−6 m31 mL×1 min60 sQ = 1.80\frac{\text{mL}}{\text{min}} \times \frac{10^{-6}\text{ m}^3}{1\text{ mL}} \times \frac{1\text{ min}}{60\text{ s}} before inserting the number into any formula.

CORRECTED PATH
ACEpath Teacher's Demonstration
  1. Convert the volumetric flow rate: Q=1.80×10−6 m360 s=3.00×10−8 m3/sQ = \frac{1.80 \times 10^{-6}\text{ m}^3}{60\text{ s}} = 3.00 \times 10^{-8}\text{ m}^3/\text{s}.

  2. Substitute the corrected QQ into the isolated equation: μ=π⋅(1.00×10−3 m)4⋅4500 Pa8⋅0.150 m⋅3.00×10−8 m3/s=4.50π×10−93.60×10−8=π8 Pa⋅s≈0.393 Pa⋅s\mu = \frac{\pi \cdot (1.00 \times 10^{-3}\text{ m})^4 \cdot 4500\text{ Pa}}{8 \cdot 0.150\text{ m} \cdot 3.00 \times 10^{-8}\text{ m}^3/\text{s}} = \frac{4.50\pi \times 10^{-9}}{3.60 \times 10^{-8}} = \frac{\pi}{8}\text{ Pa}\cdot\text{s} \approx 0.393\text{ Pa}\cdot\text{s}.

  3. Convert to centipoise: μ=0.3927 Pa⋅s×1000 cP/(Pa⋅s)=393 cP\mu = 0.3927\text{ Pa}\cdot\text{s} \times 1000\text{ cP}/(\text{Pa}\cdot\text{s}) = 393\text{ cP}.

Final answer

μ=0.393 Pa⋅s=393 cP\mu = 0.393\text{ Pa}\cdot\text{s} = 393\text{ cP}

NEXT STEPS
Final Takeaways from Your Teacher
  • Always write compound rate units as explicit fractions (e.g., mLmin\frac{\text{mL}}{\text{min}}) and convert the time denominator explicitly before numerical substitution.

  • Perform a sanity check on common liquid viscosities: light engine oils typically range between 50 cP50\text{ cP} and 500 cP500\text{ cP} at room temperature, whereas 6.54 cP6.54\text{ cP} would indicate a fluid closer to water (1 cP1\text{ cP}).

  • Practice multi-step factor-label conversions involving non-SI time and volume units.

OverviewACEpath Teacher's Diagnosis

03 The student raises the radius to the first power instead of the fourth power when evaluating Poiseuille's equation, leading to an incorrect result by many orders of magnitude.

Sample Student Work

Reviewed by ACEpath Teacher
Step-by-Step Review
Walk Through It with Your Teacher
Incorrect
  1. Step 1

    μ=π⋅R⋅ΔP8⋅L⋅Q\mu = \frac{\pi \cdot R \cdot \Delta P}{8 \cdot L \cdot Q}

    Correct

    μ=πR4ΔP8LQ\mu = \frac{\pi R^4 \Delta P}{8 L Q}

    You omitted the exponent of 44 on the radius RR when isolating μ\mu from ΔP=8μLQπR4\Delta P = \frac{8 \mu L Q}{\pi R^4}. Checking the dimensions of your expression yields m⋅Pam⋅(m3/s)=Pa⋅sm3\frac{\text{m} \cdot \text{Pa}}{\text{m} \cdot (\text{m}^3/\text{s})} = \frac{\text{Pa}\cdot\text{s}}{\text{m}^3}, rather than the required viscosity unit Pa⋅s\text{Pa}\cdot\text{s}.

  2. Step 2

    L=0.150 m  ⟹  R=1.00×10−3 m  ⟹  ΔP=4500 Pa  ⟹  Q=3.00×10−8 m3/sL = 0.150\text{ m} \implies R = 1.00 \times 10^{-3}\text{ m} \implies \Delta P = 4500\text{ Pa} \implies Q = 3.00 \times 10^{-8}\text{ m}^3/\text{s}

  3. Step 3

    μ=π⋅(1.00×10−3)⋅45008⋅0.150⋅3.00×10−8=3.93×108 Pa⋅s\mu = \frac{\pi \cdot (1.00 \times 10^{-3}) \cdot 4500}{8 \cdot 0.150 \cdot 3.00 \times 10^{-8}} = 3.93 \times 10^8\text{ Pa}\cdot\text{s}

    Correct

    μ=π⋅(1.00×10−3 m)4⋅4500 Pa8⋅0.150 m⋅3.00×10−8 m3/s=0.393 Pa⋅s\mu = \frac{\pi \cdot (1.00 \times 10^{-3}\text{ m})^4 \cdot 4500\text{ Pa}}{8 \cdot 0.150\text{ m} \cdot 3.00 \times 10^{-8}\text{ m}^3/\text{s}} = 0.393\text{ Pa}\cdot\text{s}

    Your numerical evaluation is mathematically consistent with your written formula, but because RR was raised to the first power instead of R4R^4, the result is too large by a factor of R−3=109R^{-3} = 10^9.

  4. Step 4

    μ=3.93×1011 cP\mu = 3.93 \times 10^{11}\text{ cP}

    Correct

    μ=0.393 Pa⋅s×1000 cP1 Pa⋅s=393 cP\mu = 0.393\text{ Pa}\cdot\text{s} \times \frac{1000\text{ cP}}{1\text{ Pa}\cdot\text{s}} = 393\text{ cP}

    You applied the correct unit conversion ratio (1 Pa⋅s=1000 cP1\text{ Pa}\cdot\text{s} = 1000\text{ cP}), but the final value inherits the order-of-magnitude error from Step 1.

WHAT YOU DID WELL
Compliments from Your Teacher

You correctly halved the internal diameter to determine the tube radius.

You accurately converted all measurement units into standard SI base units before computing.

You correctly applied the conversion factor between pascal-seconds and centipoise.

THE ERROR
Where Your ACEpath Teacher Spotted the Mistake
Formula transcription / missing power

You dropped the exponent 44 on radius RR when copying and rearranging Poiseuille's law.

WHY IT HAPPENED
ACEpath Teacher's Diagnosis

Skipping the dimensional analysis check requested by the problem allowed the missing power of length (R4R^4 vs. RR) to go unnoticed before numerical substitution.

REASONING PATH
Your ACEpath teacher’s suggested approach
Dimensionally validate isolated variables
  • Before substituting numbers, check that [Right-Hand Side]=[Target Variable][\text{Right-Hand Side}] = [\text{Target Variable}].

  • Verify that [R4][ΔP][L][Q]=m4⋅Pam⋅(m3/s)=Pa⋅s\frac{[R^4][\Delta P]}{[L][Q]} = \frac{\text{m}^4 \cdot \text{Pa}}{\text{m} \cdot (\text{m}^3/\text{s})} = \text{Pa}\cdot\text{s}.

Physical reasonableness check
  • Common liquid lubricants have dynamic viscosities between 10−2 Pa⋅s10^{-2}\text{ Pa}\cdot\text{s} and 101 Pa⋅s10^1\text{ Pa}\cdot\text{s} (10 cP10\text{ cP} to 104 cP10^4\text{ cP}).

  • A result of 108 Pa⋅s10^8\text{ Pa}\cdot\text{s} (greater than asphalt or mantle rock) should immediately signal a missing small-factor power such as (10−3)4(10^{-3})^4.

CORRECTED PATH
ACEpath Teacher's Demonstration
  1. Rearrange ΔP=8μLQπR4\Delta P = \frac{8 \mu L Q}{\pi R^4} to isolate μ\mu: μ=πR4ΔP8LQ\mu = \frac{\pi R^4 \Delta P}{8 L Q}.

  2. Substitute the converted SI values: μ=π(1.00×10−3 m)4(4500 Pa)8(0.150 m)(3.00×10−8 m3/s)\mu = \frac{\pi (1.00 \times 10^{-3}\text{ m})^4 (4500\text{ Pa})}{8 (0.150\text{ m}) (3.00 \times 10^{-8}\text{ m}^3/\text{s})}.

  3. Simplify the powers of ten: μ=4500π×10−123.60×10−8=4.50π×10−93.60×10−8=π8≈0.393 Pa⋅s\mu = \frac{4500\pi \times 10^{-12}}{3.60 \times 10^{-8}} = \frac{4.50\pi \times 10^{-9}}{3.60 \times 10^{-8}} = \frac{\pi}{8} \approx 0.393\text{ Pa}\cdot\text{s}.

  4. Convert to centipoise: 0.3927 Pa⋅s×1000 cP/(Pa⋅s)=393 cP0.3927\text{ Pa}\cdot\text{s} \times 1000\text{ cP}/(\text{Pa}\cdot\text{s}) = 393\text{ cP}.

Final answer

μ=0.393 Pa⋅s=393 cP\mu = 0.393\text{ Pa}\cdot\text{s} = 393\text{ cP}

NEXT STEPS
Final Takeaways from Your Teacher
  • Practice isolating variables in equations with higher powers (r3r^3, r4r^4, T2T^2) and circling exponents during algebraic rearrangement.

  • Perform a quick dimensional sanity check on the symbolic formula before plugging in numerical values.

Lock it in

Similar question

01

Water at 20∘C20^\circ\text{C} flows through a horizontal hypodermic needle of length L=4.00 cmL = 4.00\text{ cm} and internal diameter d=0.400 mmd = 0.400\text{ mm}. If the dynamic viscosity of water is μ=1.00×10−3 Pa⋅s\mu = 1.00 \times 10^{-3}\text{ Pa}\cdot\text{s} and the pressure drop is ΔP=8.00 kPa\Delta P = 8.00\text{ kPa}, calculate the volumetric flow rate QQ in mL/min\text{mL/min} using Poiseuille's law.

Answer

Q=1.88 mL/minQ = 1.88\text{ mL/min}. Key steps: R=d/2=0.200 mm=2.00×10−4 mR = d/2 = 0.200\text{ mm} = 2.00 \times 10^{-4}\text{ m}. Rearrange to Q=πR4ΔP8μL=π(2.00×10−4 m)4(8000 Pa)8(1.00×10−3 Pa⋅s)(0.0400 m)=3.142×10−8 m3/sQ = \frac{\pi R^4 \Delta P}{8 \mu L} = \frac{\pi (2.00 \times 10^{-4}\text{ m})^4 (8000\text{ Pa})}{8 (1.00 \times 10^{-3}\text{ Pa}\cdot\text{s})(0.0400\text{ m})} = 3.142 \times 10^{-8}\text{ m}^3/\text{s}. Convert: 3.142×10−8 m3/s×106 mL/m3×60 s/min≈1.88 mL/min3.142 \times 10^{-8}\text{ m}^3/\text{s} \times 10^6\text{ mL/m}^3 \times 60\text{ s/min} \approx 1.88\text{ mL/min}.

02

A blood vessel of length 2.50 cm2.50\text{ cm} experiences a pressure drop of 400 Pa400\text{ Pa} under a blood flow of Q=1.20×10−7 m3/sQ = 1.20 \times 10^{-7}\text{ m}^3/\text{s}. Assuming Poiseuille flow with blood viscosity μ=4.00×10−3 Pa⋅s\mu = 4.00 \times 10^{-3}\text{ Pa}\cdot\text{s}, find the internal diameter dd of the vessel in millimeters.

Answer

d=1.75 mmd = 1.75\text{ mm}. Key steps: Rearrange for radius: R=(8μLQπΔP)1/4=(8(4.00×10−3)(0.0250)(1.20×10−7)π(400))1/4=(7.639×10−14)1/4≈8.76×10−4 m=0.876 mmR = \left(\frac{8 \mu L Q}{\pi \Delta P}\right)^{1/4} = \left(\frac{8 (4.00 \times 10^{-3})(0.0250)(1.20 \times 10^{-7})}{\pi (400)}\right)^{1/4} = (7.639 \times 10^{-14})^{1/4} \approx 8.76 \times 10^{-4}\text{ m} = 0.876\text{ mm}. Then diameter d=2R=1.75 mmd = 2R = 1.75\text{ mm}.

03

Two horizontal pipes AA and BB have the same length and carry the same fluid under identical pressure drops. Pipe AA has diameter dA=10.0 mmd_A = 10.0\text{ mm}, and Pipe BB has diameter dB=20.0 mmd_B = 20.0\text{ mm}. If an engineer accidentally calculates the flow rate through pipe AA using dAd_A in place of RAR_A, and through pipe BB using the correct radius RBR_B, what is the ratio of their calculated flow rates QAcalc/QBQ_A^{\text{calc}} / Q_B?

Answer

QAcalc/QB=1Q_A^{\text{calc}} / Q_B = 1. Key steps: Poiseuille flow has Q∝R4Q \propto R^4. Pipe BB correct rate: QB∝RB4=(10 mm)4Q_B \propto R_B^4 = (10\text{ mm})^4. Pipe AA incorrect rate replaces RA=5 mmR_A = 5\text{ mm} with dA=10 mmd_A = 10\text{ mm}, yielding QAcalc∝(10 mm)4Q_A^{\text{calc}} \propto (10\text{ mm})^4. Thus, the ratio is (10)4/(10)4=1(10)^4 / (10)^4 = 1, demonstrating that confusing diameter with radius artificially quadrupled the effective radius, scaling flow by a factor of 1616.