In a nuclear forensic analytical lab, the activity of a liquid waste sample containing strontium-90 (3890Sr{}_{38}^{90}\text{Sr}, a pure β−\beta^- emitter with Emax⁡=0.546 MeVE_{\max} = 0.546\text{ MeV}) is determined using a calibrated liquid scintillation counter. An aliquot of 2.50 mL2.50\text{ mL} of the aqueous waste is combined with 15.0 mL15.0\text{ mL} of scintillation cocktail. Over a 20.0-minute20.0\text{-minute} counting interval, the instrument records a total of 58420 counts58420\text{ counts}. A blank background sample counted under identical geometry for 60.0 minutes60.0\text{ minutes} registers 1440 counts1440\text{ counts}. If the absolute counting efficiency of the detector for 90Sr{}^{90}\text{Sr} is 86.5%86.5\%, calculate the corrected net count rate in counts per minute (cpm\text{cpm}), and determine the volumetric activity of 90Sr{}^{90}\text{Sr} in the original waste liquid in kBq/L\text{kBq/L}.

Answer and steps

Final answer

Net count rate =2.897×103 cpm= 2.897 \times 10^3\text{ cpm}; Volumetric activity =22.3 kBq/L= 22.3\text{ kBq/L}

Steps

01

Calculate the gross and background count rates

The instrument records raw counts over specified time intervals. First, convert each total count into a count rate in counts per minute (cpm\text{cpm}):

Gross count rate (Rgross)=Gross countstgross=58420 counts20.0 min=2921 cpm\text{Gross count rate } (R_{\text{gross}}) = \frac{\text{Gross counts}}{t_{\text{gross}}} = \frac{58420\text{ counts}}{20.0\text{ min}} = 2921\text{ cpm}

Background count rate (Rbg)=Background countstbg=1440 counts60.0 min=24.0 cpm\text{Background count rate } (R_{\text{bg}}) = \frac{\text{Background counts}}{t_{\text{bg}}} = \frac{1440\text{ counts}}{60.0\text{ min}} = 24.0\text{ cpm}


02

Determine the net count rate (RnetR_{\text{net}})

Subtract the background rate from the gross count rate to find the net count rate produced solely by the sample:

Rnet=Rgross−Rbg=2921 cpm−24.0 cpm=2897 cpm=2.897×103 cpmR_{\text{net}} = R_{\text{gross}} - R_{\text{bg}} = 2921\text{ cpm} - 24.0\text{ cpm} = 2897\text{ cpm} = 2.897 \times 10^3\text{ cpm}


03

Convert the net count rate to true nuclear activity (AA)

The counter detects a fraction of all radioactive disintegrations according to its absolute counting efficiency (ε=86.5%=0.865\varepsilon = 86.5\% = 0.865). The true rate of disintegrations per minute (dpm\text{dpm}) in the aliquot is:

Disintegration rate (dpm)=Rnetε=2897 cpm0.865≈3349.13 dpm\text{Disintegration rate } (\text{dpm}) = \frac{R_{\text{net}}}{\varepsilon} = \frac{2897\text{ cpm}}{0.865} \approx 3349.13\text{ dpm}

One Becquerel (Bq\text{Bq}) is defined as one nuclear disintegration per second (1 Bq=1 dps=60 dpm1\text{ Bq} = 1\text{ dps} = 60\text{ dpm}):

Aaliquot=3349.13 dpm60 s/min≈55.82 BqA_{\text{aliquot}} = \frac{3349.13\text{ dpm}}{60\text{ s/min}} \approx 55.82\text{ Bq}

Common Pitfall: Forgetting to divide dpm\text{dpm} by 6060 before calculating Becquerels leads to an activity value that is sixty times too large.


04

Calculate the volumetric activity in the original waste

The aliquot measured has a volume of Valiquot=2.50 mL=2.50×10−3 LV_{\text{aliquot}} = 2.50\text{ mL} = 2.50 \times 10^{-3}\text{ L}. The volumetric activity (CAC_A) of X90X2290Sr\ce{^{90}Sr} in the liquid waste is:

CA=AaliquotValiquot=55.82 Bq2.50×10−3 L=22328 Bq/L=22.3 kBq/LC_A = \frac{A_{\text{aliquot}}}{V_{\text{aliquot}}} = \frac{55.82\text{ Bq}}{2.50 \times 10^{-3}\text{ L}} = 22328\text{ Bq/L} = 22.3\text{ kBq/L}

Typical errors

OverviewACEpath Teacher's Diagnosis

01 The student uses the total cocktail mixture volume of 17.50 mL instead of the original 2.50 mL aliquot volume when calculating volumetric activity.

Sample Student Work

Reviewed by ACEpath Teacher
Step-by-Step Review
Walk Through It with Your Teacher
Partially correct
  1. Step 1

    Rgross=58420/20.0=2921 cpm  ⟹  Rbkg=1440/60.0=24 cpm  ⟹  Rnet=2921−24=2897 cpmR_{\text{gross}} = 58420 / 20.0 = 2921\text{ cpm} \implies R_{\text{bkg}} = 1440 / 60.0 = 24\text{ cpm} \implies R_{\text{net}} = 2921 - 24 = 2897\text{ cpm}

  2. Step 2

    Aaliquot=2897/(0.865×60)=55.82 BqA_{\text{aliquot}} = 2897 / (0.865 \times 60) = 55.82\text{ Bq}

  3. Step 3

    Vtotal=2.50 mL+15.0 mL=17.50 mL  ⟹  Activity concentration=55.82 Bq/17.50 mL=3.19 kBq/LV_{\text{total}} = 2.50\text{ mL} + 15.0\text{ mL} = 17.50\text{ mL} \implies \text{Activity concentration} = 55.82\text{ Bq} / 17.50\text{ mL} = 3.19\text{ kBq/L}

    Correct

    CA=AaliquotValiquot=55.82 Bq2.50×10−3 L=22.3 kBq/LC_A = \frac{A_{\text{aliquot}}}{V_{\text{aliquot}}} = \frac{55.82\text{ Bq}}{2.50 \times 10^{-3}\text{ L}} = 22.3\text{ kBq/L}

    You divided by the combined volume of the waste and scintillation cocktail (17.50 mL17.50\text{ mL}). The question asked for the volumetric activity of the original waste liquid, so you should divide only by the aliquot volume of waste introduced (2.50 mL2.50\text{ mL}).

WHAT YOU DID WELL
Compliments from Your Teacher

You calculated both the gross and background count rates accurately from the total counts and counting times.

You subtracted background counts correctly to determine the net count rate of 2897 cpm2897\text{ cpm}.

You correctly used the counting efficiency and time conversion factor to compute the sample activity in Becquerels.

THE ERROR
Where Your ACEpath Teacher Spotted the Mistake
Incorrect reference volume in concentration calculation

no

WHY IT HAPPENED
ACEpath Teacher's Diagnosis

You treated the scintillation cocktail as a diluent that defines the final sample volume of interest, rather than recognizing that the question asked for the concentration in the original undiluted liquid waste.

REASONING PATH
Your ACEpath teacher’s suggested approach
Distinguish Analytical Aliquot from Vial Matrix
  • Scintillation cocktail acts only as an energy-transfer detection medium; it does not change the amount of analyte drawn from the original sample source.

  • Always divide the measured activity by the volume of original sample taken (ValiquotV_{\text{aliquot}}), not the total vial volume (Valiquot+VcocktailV_{\text{aliquot}} + V_{\text{cocktail}}).

CORRECTED PATH
ACEpath Teacher's Demonstration
  1. Identify the aliquot volume of original waste liquid: Valiquot=2.50 mL=2.50×10−3 LV_{\text{aliquot}} = 2.50\text{ mL} = 2.50 \times 10^{-3}\text{ L}.

  2. Divide the measured aliquot activity by this volume: CA=55.82 Bq2.50×10−3 L=22328 Bq/LC_A = \frac{55.82\text{ Bq}}{2.50 \times 10^{-3}\text{ L}} = 22328\text{ Bq/L}.

  3. Convert to the requested units: 22328 Bq/L=22.3 kBq/L22328\text{ Bq/L} = 22.3\text{ kBq/L}.

Final answer

Net count rate =2.897×103 cpm= 2.897 \times 10^3\text{ cpm}; Volumetric activity =22.3 kBq/L= 22.3\text{ kBq/L}

NEXT STEPS
Final Takeaways from Your Teacher
  • Review the distinction between sample aliquot volume and total counting vial volume in scintillation counting.

  • Practice identifying the target reference solution when computing concentrations in radiochemistry and spectroscopy dilution problems.

OverviewACEpath Teacher's Diagnosis

02 The student forgets to divide disintegrations per minute by 60 to convert to Becquerels, treating 1 cpm/efficiency directly as 1 Bq.

Sample Student Work

Reviewed by ACEpath Teacher
Step-by-Step Review
Walk Through It with Your Teacher
Partially correct
  1. Step 1

    Rgross=5842020.0=2921 cpm  ⟹  Rbkg=144060.0=24 cpmR_{\text{gross}} = \frac{58420}{20.0} = 2921\text{ cpm} \implies R_{\text{bkg}} = \frac{1440}{60.0} = 24\text{ cpm}

  2. Step 2

    Rnet=2921−24=2897 cpmR_{\text{net}} = 2921 - 24 = 2897\text{ cpm}

  3. Step 3

    Disintegration rate=28970.865=3349 Bq\text{Disintegration rate} = \frac{2897}{0.865} = 3349\text{ Bq}

    Correct

    A=Rnetε×60 s/min=2897 cpm0.865×60 s/min≈55.82 BqA = \frac{R_{\text{net}}}{\varepsilon \times 60\text{ s/min}} = \frac{2897\text{ cpm}}{0.865 \times 60\text{ s/min}} \approx 55.82\text{ Bq}

    Dividing counts per minute by the efficiency gives disintegrations per minute (dpm\text{dpm}), not Becquerels (Bq\text{Bq}). Because 1 Bq=1 disintegration per second=60 dpm1\text{ Bq} = 1\text{ disintegration per second} = 60\text{ dpm}, you need to divide by 6060 to convert to Bq\text{Bq}.

  4. Step 4

    Volumetric activity=3349 Bq2.50 mL=1339.6 kBq/L=1.34×103 kBq/L\text{Volumetric activity} = \frac{3349\text{ Bq}}{2.50\text{ mL}} = 1339.6\text{ kBq/L} = 1.34 \times 10^3\text{ kBq/L}

    Correct

    CA=55.82 Bq2.50×10−3 L=2.23×104 Bq/L=22.3 kBq/LC_A = \frac{55.82\text{ Bq}}{2.50 \times 10^{-3}\text{ L}} = 2.23 \times 10^4\text{ Bq/L} = 22.3\text{ kBq/L}

    Your concentration calculation and unit prefix conversion were set up properly, but the numerical result is 6060 times higher due to the prior dpm\text{dpm}-to-Bq\text{Bq} unit error.

WHAT YOU DID WELL
Compliments from Your Teacher

You correctly calculated both the gross and background count rates per minute.

You accurately found the net count rate by subtracting background from gross.

You correctly divided by the fractional counting efficiency (0.8650.865) to account for detector efficiency.

You used the correct aliquot volume and properly converted Bq/mL\text{Bq/mL} to kBq/L\text{kBq/L}.

THE ERROR
Where Your ACEpath Teacher Spotted the Mistake
Time unit conversion in radioactivity (dpm to Bq)

You labeled the result of Rnet/εR_{\text{net}}/\varepsilon directly as Becquerels (Bq\text{Bq}) rather than disintegrations per minute (dpm\text{dpm}).

WHY IT HAPPENED
ACEpath Teacher's Diagnosis

Overlooking the definition of the SI unit of activity: 1 Bq=1 disintegration per second (dps)1\text{ Bq} = 1\text{ disintegration per second}\text{ (dps)}. When counting rates are recorded in cpm\text{cpm}, dividing by detector efficiency yields dpm\text{dpm}, which must be divided by 60 s/min60\text{ s/min} to yield Bq\text{Bq}.

REASONING PATH
Your ACEpath teacher’s suggested approach
Unit Alignment: cpm to dpm to Bq
  • dpm=cpmε\text{dpm} = \frac{\text{cpm}}{\varepsilon}, and 1 Bq=1 dps=1 dpm601\text{ Bq} = 1\text{ dps} = \frac{1\text{ dpm}}{60}. Therefore, A (Bq)=dpm60=cpm60×εA\text{ (Bq)} = \frac{\text{dpm}}{60} = \frac{\text{cpm}}{60 \times \varepsilon}.

  • Always write explicit time units (counts/min\text{counts/min} and disintegrations/min\text{disintegrations/min}) and include the conversion factor 1 min60 s\frac{1\text{ min}}{60\text{ s}} before introducing Bq\text{Bq}.

CORRECTED PATH
ACEpath Teacher's Demonstration
  1. Convert disintegrations per minute to Becquerels: A=3349.13 dpm60 s/min≈55.82 BqA = \frac{3349.13\text{ dpm}}{60\text{ s/min}} \approx 55.82\text{ Bq}.

  2. Calculate the volumetric activity of the 2.50 mL2.50\text{ mL} aliquot: CA=55.82 Bq2.50×10−3 L=22328 Bq/L=22.3 kBq/LC_A = \frac{55.82\text{ Bq}}{2.50 \times 10^{-3}\text{ L}} = 22328\text{ Bq/L} = 22.3\text{ kBq/L}.

Final answer

Net count rate =2.897×103 cpm= 2.897 \times 10^3\text{ cpm}; Volumetric activity =22.3 kBq/L= 22.3\text{ kBq/L}

NEXT STEPS
Final Takeaways from Your Teacher
  • Review the distinction between counting units (cpm, dpm) and SI activity units (Bq = dps, Ci = 3.7×1010 dps3.7 \times 10^{10}\text{ dps}).

  • Practice converting net counts per minute directly into Becquerels using the unified relationship A (Bq)=Rnet60×εA\text{ (Bq)} = \frac{R_{\text{net}}}{60 \times \varepsilon}.

OverviewACEpath Teacher's Diagnosis

03 The student subtracts raw counts directly without dividing by their respective counting intervals to find count rates.

Sample Student Work

Reviewed by ACEpath Teacher
Step-by-Step Review
Walk Through It with Your Teacher
Incorrect
  1. Step 1

    Net counts=58420−1440=56980 counts  ⟹  Net count rate=56980/20.0=2849 cpm\text{Net counts} = 58420 - 1440 = 56980\text{ counts} \implies \text{Net count rate} = 56980 / 20.0 = 2849\text{ cpm}

    Correct

    Rgross=58420 counts20.0 min=2921 cpm,Rbg=1440 counts60.0 min=24.0 cpm  ⟹  Rnet=2921−24.0=2897 cpmR_{\text{gross}} = \frac{58420\text{ counts}}{20.0\text{ min}} = 2921\text{ cpm},\quad R_{\text{bg}} = \frac{1440\text{ counts}}{60.0\text{ min}} = 24.0\text{ cpm} \implies R_{\text{net}} = 2921 - 24.0 = 2897\text{ cpm}

    You subtracted raw background counts from raw gross counts directly, but the two counts were acquired over different time periods (20.0 min20.0\text{ min} vs 60.0 min60.0\text{ min}). You need to determine the count rate for each measurement first (R=N/tR = N / t) and then calculate Rnet=Rgross−RbgR_{\text{net}} = R_{\text{gross}} - R_{\text{bg}}.

  2. Step 2

    Aaliquot=2849/(0.865×60)=54.89 BqA_{\text{aliquot}} = 2849 / (0.865 \times 60) = 54.89\text{ Bq}

    Correct

    Aaliquot=2897 cpm0.865×60 s/min≈55.82 BqA_{\text{aliquot}} = \frac{2897\text{ cpm}}{0.865 \times 60\text{ s/min}} \approx 55.82\text{ Bq}

    Your formula correctly divides the count rate by efficiency and 60 s/min60\text{ s/min} to convert cpm\text{cpm} to Bq\text{Bq}, but the numerical result is affected by the incorrect net count rate from Step 1.

  3. Step 3

    Volumetric activity=54.89 Bq/2.50 mL=22.0 kBq/L\text{Volumetric activity} = 54.89\text{ Bq} / 2.50\text{ mL} = 22.0\text{ kBq/L}

    Correct

    CA=55.82 Bq2.50×10−3 L≈22.3 kBq/LC_A = \frac{55.82\text{ Bq}}{2.50 \times 10^{-3}\text{ L}} \approx 22.3\text{ kBq/L}

    Your volume conversion and scaling from Bq/mL\text{Bq/mL} to kBq/L\text{kBq/L} are executed correctly, but this final value inherits the discrepancy from the earlier steps.

WHAT YOU DID WELL
Compliments from Your Teacher

You correctly recognized that absolute activity requires dividing net count rate by efficiency.

You handled the unit conversion between disintegrations per minute (dpm\text{dpm}) and Becquerels (Bq\text{Bq}) by including the factor of 60 s/min60\text{ s/min}.

You correctly converted the aliquot activity into the volumetric activity of the bulk liquid waste in kBq/L\text{kBq/L}.

THE ERROR
Where Your ACEpath Teacher Spotted the Mistake
Unequal Counting Times Subtraction

You subtracted total background counts from total gross counts even though their acquisition durations were different.

WHY IT HAPPENED
ACEpath Teacher's Diagnosis

Conflating total accumulated counts (NN) with count rates (R=N/tR = N/t). When counting intervals (t1t_1 and t2t_2) are unequal, raw counts represent different time bases and are not directly additive or subtractable.

REASONING PATH
Your ACEpath teacher’s suggested approach
Time Normalization Check
  • Always convert gross and background raw counts to rates (cpm\text{cpm} or cps\text{cps}) before taking differences: Rnet=Ngrosstgross−NbgtbgR_{\text{net}} = \frac{N_{\text{gross}}}{t_{\text{gross}}} - \frac{N_{\text{bg}}}{t_{\text{bg}}}.

  • Check whether the problem specifies identical or different counting times for the sample and background before writing your first subtraction line.

CORRECTED PATH
ACEpath Teacher's Demonstration
  1. Calculate the individual count rates: Rgross=5842020.0=2921 cpmR_{\text{gross}} = \frac{58420}{20.0} = 2921\text{ cpm} and Rbg=144060.0=24.0 cpmR_{\text{bg}} = \frac{1440}{60.0} = 24.0\text{ cpm}.

  2. Subtract the background rate from the gross rate: Rnet=2921−24.0=2897 cpm=2.897×103 cpmR_{\text{net}} = 2921 - 24.0 = 2897\text{ cpm} = 2.897 \times 10^3\text{ cpm}.

  3. Convert the net count rate to activity in Bq\text{Bq}: Aaliquot=28970.865×60≈55.82 BqA_{\text{aliquot}} = \frac{2897}{0.865 \times 60} \approx 55.82\text{ Bq}.

  4. Divide by the aliquot volume to find the volumetric activity: CA=55.82 Bq2.50×10−3 L=22328 Bq/L≈22.3 kBq/LC_A = \frac{55.82\text{ Bq}}{2.50 \times 10^{-3}\text{ L}} = 22328\text{ Bq/L} \approx 22.3\text{ kBq/L}.

Final answer

Net count rate =2.897×103 cpm= 2.897 \times 10^3\text{ cpm}; Volumetric activity =22.3 kBq/L= 22.3\text{ kBq/L}

NEXT STEPS
Final Takeaways from Your Teacher
  • Review radiation measurement problems where background counting times are longer than sample counting times to practice normalizing counts to rates first.

  • Re-solve this problem independently from scratch, ensuring RgrossR_{\text{gross}} and RbgR_{\text{bg}} are calculated as distinct intermediate values.

Lock it in

Similar question

01

A 5.00 mL5.00\text{ mL} aliquot of tritiated groundwater is mixed with 15.0 mL15.0\text{ mL} of scintillation fluid. The detector records a net count rate of 1800 cpm1800\text{ cpm} with an efficiency of 40.0%40.0\%. Calculate the tritium concentration in the groundwater in Bq/L\text{Bq/L}.

Answer
  1. True activity: A=1800 cpm0.400×60 s/min=75.0 BqA = \frac{1800\text{ cpm}}{0.400 \times 60\text{ s/min}} = 75.0\text{ Bq}.\n2. Volumetric activity: CA=75.0 Bq5.00×10−3 L=1.50×104 Bq/L=15.0 kBq/LC_A = \frac{75.0\text{ Bq}}{5.00 \times 10^{-3}\text{ L}} = 1.50 \times 10^4\text{ Bq/L} = 15.0\text{ kBq/L}.
02

A 0.500 mL0.500\text{ mL} bioassay urine sample is added to 10.0 mL10.0\text{ mL} of cocktail. Gross counting over 10.0 min10.0\text{ min} gives 3650 counts3650\text{ counts}, while background over 30.0 min30.0\text{ min} gives 450 counts450\text{ counts}. With an efficiency of 70.0%70.0\%, find the volumetric activity of the urine in Bq/mL\text{Bq/mL}.

Answer
  1. Count rates: Rgross=365 cpmR_{\text{gross}} = 365\text{ cpm}, Rbg=15 cpm  ⟹  Rnet=350 cpmR_{\text{bg}} = 15\text{ cpm} \implies R_{\text{net}} = 350\text{ cpm}.\n2. Aliquot activity: A=3500.700×60=8.33 BqA = \frac{350}{0.700 \times 60} = 8.33\text{ Bq}.\n3. Volumetric activity: CA=8.33 Bq0.500 mL=16.7 Bq/mLC_A = \frac{8.33\text{ Bq}}{0.500\text{ mL}} = 16.7\text{ Bq/mL}.
03

An environmental soil leachate is diluted 1:101:10 with deionized water. A 2.00 mL2.00\text{ mL} aliquot of this diluted solution is combined with 18.0 mL18.0\text{ mL} of scintillation cocktail, yielding a measured activity of 48.0 Bq48.0\text{ Bq}. Determine the activity concentration of the undiluted leachate in kBq/L\text{kBq/L}.

Answer
  1. Concentration in diluted solution: Cdil=48.0 Bq2.00×10−3 L=24.0 kBq/LC_{\text{dil}} = \frac{48.0\text{ Bq}}{2.00 \times 10^{-3}\text{ L}} = 24.0\text{ kBq/L}.\n2. Account for 1:101:10 dilution: Cundiluted=24.0 kBq/L×10=240 kBq/LC_{\text{undiluted}} = 24.0\text{ kBq/L} \times 10 = 240\text{ kBq/L}.