A chemistry student standardizes an unknown aqueous solution of sodium hydroxide (NaOH\text{NaOH}) by titrating three separate 0.5106 g0.5106\text{ g} samples of primary-standard potassium hydrogen phthalate (KHP\text{KHP}, KHC8H4O4\text{KHC}_8\text{H}_4\text{O}_4, molar mass =204.22 g⋅mol−1= 204.22\text{ g}\cdot\text{mol}^{-1}) dissolved in 50.0 mL50.0\text{ mL} of deionized water. The titration reaction is KHC8H4O4(aq)+NaOH(aq)→KNaC8H4O4(aq)+H2O(l)\text{KHC}_8\text{H}_4\text{O}_4(aq) + \text{NaOH}(aq) \rightarrow \text{KNaC}_8\text{H}_4\text{O}_4(aq) + \text{H}_2\text{O}(l). The student records the following initial and final burette volumes: Trial 1: 0.50 mL0.50\text{ mL} to 25.62 mL25.62\text{ mL}; Trial 2: 1.15 mL1.15\text{ mL} to 26.23 mL26.23\text{ mL}; Trial 3: 3.40 mL3.40\text{ mL} to 28.54 mL28.54\text{ mL}. Prior to Trial 3, the student noticed a persistent air bubble trapped in the burette tip that dislodged during the delivery. Evaluate whether Trial 3 should be retained or discarded based on experimental systematic error, calculate the mean concentration of NaOH\text{NaOH} from the valid trials, and determine whether the experimental result agrees with the manufacturer's nominal claim of 0.1000 M±0.0010 M0.1000\text{ M} \pm 0.0010\text{ M}.

Answer and steps

Final answer

Titer volumes delivered: Trial 1: 25.62−0.50=25.12 mL25.62 - 0.50 = 25.12\text{ mL}; Trial 2: 26.23−1.15=25.08 mL26.23 - 1.15 = 25.08\text{ mL}; Trial 3: 28.54−3.40=25.14 mL28.54 - 3.40 = 25.14\text{ mL}. An air bubble filling with liquid during titration consumes apparent burette volume without delivering titrant to the flask, causing an artificially inflated volume and an artificially low calculated concentration. However, here V3=25.14 mLV_3 = 25.14\text{ mL} is within normal random reading variation (0.04 mL0.04\text{ mL} of the mean of Trials 1 and 2), indicating the bubble was purged during flushing/priming rather than during titrant delivery, or had a negligible volume (< 0.05 mL0.05\text{ mL}). Moles of KHP per trial: 0.5106 g204.22 g⋅mol−1=2.5002×10−3 mol\frac{0.5106\text{ g}}{204.22\text{ g}\cdot\text{mol}^{-1}} = 2.5002 \times 10^{-3}\text{ mol}. Concentrations: C1=2.5002×10−3 mol0.02512 L=0.09953 MC_1 = \frac{2.5002 \times 10^{-3}\text{ mol}}{0.02512\text{ L}} = 0.09953\text{ M}; C2=2.5002×10−3 mol0.02508 L=0.09969 MC_2 = \frac{2.5002 \times 10^{-3}\text{ mol}}{0.02508\text{ L}} = 0.09969\text{ M}; C3=2.5002×10−3 mol0.02514 L=0.09945 MC_3 = \frac{2.5002 \times 10^{-3}\text{ mol}}{0.02514\text{ L}} = 0.09945\text{ M}. The average concentration across all trials is 0.09956 M0.09956\text{ M} (or across Trials 1 and 2 is 0.09961 M0.09961\text{ M}). Both fall squarely within the manufacturer's tolerance interval of [0.0990 M,0.1010 M][0.0990\text{ M}, 0.1010\text{ M}], confirming the manufacturer's claim.

Steps

01

Calculate the Delivered Volume for Each Trial

The volume of titrant (NaOH\ce{NaOH}) delivered in each run is the difference between the final and initial burette readings:

  • Trial 1:
    V1=25.62 mL−0.50 mL=25.12 mLV_1 = 25.62\text{ mL} - 0.50\text{ mL} = 25.12\text{ mL}
  • Trial 2:
    V2=26.23 mL−1.15 mL=25.08 mLV_2 = 26.23\text{ mL} - 1.15\text{ mL} = 25.08\text{ mL}
  • Trial 3:
    V3=28.54 mL−3.40 mL=25.14 mLV_3 = 28.54\text{ mL} - 3.40\text{ mL} = 25.14\text{ mL}

02

Evaluate the Air Bubble in Trial 3

  • Theoretical Impact: If an air bubble trapped below the stopcock dislodges and fills with liquid during the titration delivery, liquid from the calibrated barrel enters the tip to displace the air rather than entering the receiving flask. This records an artificially large volume (ΔV\Delta V), which leads to an artificially low calculated molarity (M=nVM = \frac{n}{V}).
  • Data Analysis: The delivered volume in Trial 3 is 25.14 mL25.14\text{ mL}. Compared to Trials 1 and 2 (mean =25.12+25.082=25.10 mL= \frac{25.12 + 25.08}{2} = 25.10\text{ mL}), the difference is only +0.04 mL+0.04\text{ mL}. This deviation lies within the typical uncertainty of a Class A 50 mL50\text{ mL} burette (±0.05 mL\pm 0.05\text{ mL}) and reading errors (±0.02 mL\pm 0.02\text{ mL} per reading).
  • Conclusion: The bubble did not introduce a significant systematic error (it either dislodged during the pre-titration zeroing/priming step or had an imperceptible volume <0.05 mL< 0.05\text{ mL}). Therefore, the data can be evaluated either with all three trials retained or by conservatively averaging only Trials 1 and 2.

03

Determine the Moles of Primary Standard (KHP\ce{KHP})

Using the molar mass of potassium hydrogen phthalate (KHCX8HX4OX4\ce{KHC8H4O4}, 204.22 g⋅mol−1204.22\text{ g}\cdot\text{mol}^{-1}):

nKHP=0.5106 g204.22 g⋅mol−1=2.5002×10−3 moln_{\ce{KHP}} = \frac{0.5106\text{ g}}{204.22\text{ g}\cdot\text{mol}^{-1}} = 2.5002 \times 10^{-3}\text{ mol}

From the balanced equation: KHCX8HX4OX4(aq)+NaOH(aq)→KNaCX8HX4OX4(aq)+HX2O(l)\ce{KHC8H4O4(aq) + NaOH(aq) -> KNaC8H4O4(aq) + H2O(l)}

The stoichiometric ratio is 1:11:1, so at the equivalence point: nNaOH=nKHP=2.5002×10−3 moln_{\ce{NaOH}} = n_{\ce{KHP}} = 2.5002 \times 10^{-3}\text{ mol}


04

Calculate the Concentration of NaOH\ce{NaOH}

Calculate the molarity for each trial using C=nV (in L)C = \frac{n}{V\text{ (in L)}}:

  • Trial 1:
    C1=2.5002×10−3 mol0.02512 L=0.09953 MC_1 = \frac{2.5002 \times 10^{-3}\text{ mol}}{0.02512\text{ L}} = 0.09953\text{ M}
  • Trial 2:
    C2=2.5002×10−3 mol0.02508 L=0.09969 MC_2 = \frac{2.5002 \times 10^{-3}\text{ mol}}{0.02508\text{ L}} = 0.09969\text{ M}
  • Trial 3:
    C3=2.5002×10−3 mol0.02514 L=0.09945 MC_3 = \frac{2.5002 \times 10^{-3}\text{ mol}}{0.02514\text{ L}} = 0.09945\text{ M}

Mean Concentration:

  • Averaging the strictly verified Trials 1 and 2: C‾1−2=0.09953 M+0.09969 M2=0.09961 M\overline{C}_{1-2} = \frac{0.09953\text{ M} + 0.09969\text{ M}}{2} = 0.09961\text{ M}
  • Averaging all three trials (Trials 1, 2, and 3): C‾all=0.09953 M+0.09969 M+0.09945 M3=0.09956 M\overline{C}_{\text{all}} = \frac{0.09953\text{ M} + 0.09969\text{ M} + 0.09945\text{ M}}{3} = 0.09956\text{ M}

05

Compare with Manufacturer Specification

The manufacturer claims a nominal concentration of 0.1000 M±0.0010 M0.1000\text{ M} \pm 0.0010\text{ M}, defining an acceptable range of: [0.0990 M,0.1010 M][0.0990\text{ M}, 0.1010\text{ M}]

Both calculated mean values (0.09961 M0.09961\text{ M} and 0.09956 M0.09956\text{ M}) lie well within the interval [0.0990 M,0.1010 M][0.0990\text{ M}, 0.1010\text{ M}], confirming that the experimental concentration agrees with the manufacturer's specification.

Typical errors

OverviewACEpath Teacher's Diagnosis

01 The student includes Trial 3 in the average despite the systematic error from the dislodged air bubble, yielding an erroneously low mean concentration.

Sample Student Work

Reviewed by ACEpath Teacher
Step-by-Step Review
Walk Through It with Your Teacher
Partially correct
  1. Step 1

    Trial 3 is retained because averaging all three trials minimizes experimental uncertainty.

    Correct

    A dislodged air bubble in the burette tip is a systematic error that artificially inflates the delivered volume (VV) and lowers the calculated molarity (M=nVM = \frac{n}{V}). Comparing V3=25.14 mLV_3 = 25.14\text{ mL} to the mean of Trials 1 and 2 (25.10 mL25.10\text{ mL}), the difference is only +0.04 mL+0.04\text{ mL}, which falls within Class A burette reading tolerance (±0.05 mL\pm 0.05\text{ mL}). Thus, the bubble volume was negligible, and Trial 3 may be retained (or conservatively discarded due to operator error).

    You justified retaining Trial 3 by stating that averaging minimizes experimental uncertainty. While averaging replicates reduces random (indeterminate) error, it does not correct or eliminate systematic (determinate) error. To evaluate Trial 3 properly, you need to analyze the physical direction of the error caused by the bubble and compare the delivered volume against the concordant trials.

  2. Step 2

    V1=25.12 mL  ⟹  V2=26.23 mL−1.15 mL=25.08 mL  ⟹  V3=28.54 mL−3.40 mL=25.14 mLV_1 = 25.12\text{ mL} \implies V_2 = 26.23\text{ mL} - 1.15\text{ mL} = 25.08\text{ mL} \implies V_3 = 28.54\text{ mL} - 3.40\text{ mL} = 25.14\text{ mL}

  3. Step 3

    n(KHP)=0.5106 g/204.22 g/mol=0.002500 mol  ⟹  M1=0.09952 M,M2=0.09968 M,M3=0.09944 Mn(\text{KHP}) = 0.5106\text{ g} / 204.22\text{ g/mol} = 0.002500\text{ mol} \implies M_1 = 0.09952\text{ M}, M_2 = 0.09968\text{ M}, M_3 = 0.09944\text{ M}

  4. Step 4

    Mean M=(0.09952+0.09968+0.09944)/3=0.09955 M\text{Mean } M = (0.09952 + 0.09968 + 0.09944)/3 = 0.09955\text{ M}

    Correct

    M‾1−2=0.09953 M+0.09969 M2=0.09961 M\overline{M}_{1-2} = \frac{0.09953\text{ M} + 0.09969\text{ M}}{2} = 0.09961\text{ M} (or M‾all=0.09956 M\overline{M}_{\text{all}} = 0.09956\text{ M} with explicit justification)

    Your arithmetic is correct, but your inclusion of Trial 3 in the average was based on an invalid rationale regarding systematic error.

  5. Step 5

    The nominal range is 0.0990 M0.0990\text{ M} to 0.1010 M0.1010\text{ M}. Since 0.09955 M0.09955\text{ M} falls within this range, the result agrees with the manufacturer's claim.

WHAT YOU DID WELL
Compliments from Your Teacher

You calculated all delivered titrant volumes accurately from the initial and final burette marks.

You determined the moles of KHP and molarity for each individual trial correctly.

You established the nominal tolerance interval and correctly judged agreement with the manufacturer's specification.

THE ERROR
Where Your ACEpath Teacher Spotted the Mistake
Systematic vs. Random Error Misconception

You treated an identified systematic error as random noise that could be reduced by simple averaging.

WHY IT HAPPENED
ACEpath Teacher's Diagnosis

You conflated random error (indeterminate fluctuations that cancel out over multiple runs) with systematic error (determinate bias that pushes measurements in one direction). When an air bubble dislodges from a burette tip during titration, liquid fills the void, falsely registering as delivered volume and biasing the concentration downward. An identified systematic error cannot be resolved by averaging; it must be evaluated by assessing its physical mechanism and comparing the trial against concordant data within known instrument tolerance.

REASONING PATH
Your ACEpath teacher’s suggested approach
Random vs. Systematic Error
  • Averaging repeated trials minimizes random uncertainty, but it never eliminates systematic bias.

  • Whenever a specific operational issue is observed (such as an air bubble, leak, or spilled analyte), treat it as a potential systematic error, not random variance.

Burette Tip Air Bubble Mechanism
  • An air bubble escaping the tip causes titrant to fill the tip instead of entering the analyte flask, making VdeliveredV_{\text{delivered}} appear too large and M=nVM = \frac{n}{V} too small.

  • Assess whether the observed titer deviation matches the predicted direction and exceeds typical burette uncertainty (±0.05 mL\pm 0.05\text{ mL}).

CORRECTED PATH
ACEpath Teacher's Demonstration
  1. Identify the systematic effect: Dislodging an air bubble draws liquid into the tip below the stopcock. This consumes registered burette volume without adding NaOH\text{NaOH} to the flask, causing Vmeasured>VtrueV_{\text{measured}} > V_{\text{true}} and Mcalc<MtrueM_{\text{calc}} < M_{\text{true}}.

  2. Compare titer values: V1=25.12 mLV_1 = 25.12\text{ mL}, V2=25.08 mLV_2 = 25.08\text{ mL}, and V3=25.14 mLV_3 = 25.14\text{ mL}. The difference between V3V_3 and the average of the clean trials (25.10 mL25.10\text{ mL}) is only +0.04 mL+0.04\text{ mL}.

  3. Decide retention based on data: Because +0.04 mL+0.04\text{ mL} is within the standard Class A reading tolerance (±0.05 mL\pm 0.05\text{ mL}), the bubble volume was negligible and Trial 3 may be kept, or conservatively discarded to report M‾1−2=0.09961 M\overline{M}_{1-2} = 0.09961\text{ M}.

  4. State the conclusion: In either case (M‾=0.09961 M\overline{M} = 0.09961\text{ M} or 0.09956 M0.09956\text{ M}), the value lies well within [0.0990 M,0.1010 M][0.0990\text{ M}, 0.1010\text{ M}], validating the manufacturer's claim.

Final answer

Trial 3 has a negligible systematic error (+0.04 mL+0.04\text{ mL}, within burette reading tolerance ±0.05 mL\pm 0.05\text{ mL}); the mean concentration is 0.09961 M0.09961\text{ M} (Trials 1–2) or 0.09956 M0.09956\text{ M} (all trials), which confirms the manufacturer's claim (0.1000 M±0.0010 M0.1000\text{ M} \pm 0.0010\text{ M}).

NEXT STEPS
Final Takeaways from Your Teacher
  • Review the distinction between random (indeterminate) error and systematic (determinate) error in quantitative volumetric analysis.

  • Practice tracing the directional impact of common laboratory errors (e.g., air bubbles, water left in burette, rinsing errors) on the final calculated concentration.

  • Use tolerance limits (such as ±0.05 mL\pm 0.05\text{ mL} for Class A 50 mL50\text{ mL} burettes) when justifying whether to retain or reject an atypical trial.

OverviewACEpath Teacher's Diagnosis

02 The student divides the moles of KHP by the volume of water used to dissolve it rather than the delivered burette volume of NaOH.

Sample Student Work

Reviewed by ACEpath Teacher
Step-by-Step Review
Walk Through It with Your Teacher
Incorrect
  1. Step 1

    Trial 3 should be discarded because the dislodged air bubble leads to an overestimate of delivered titrant.

    Correct

    Trial 1: 25.62−0.50=25.12 mL25.62 - 0.50 = 25.12\text{ mL}, Trial 2: 26.23−1.15=25.08 mL26.23 - 1.15 = 25.08\text{ mL}, Trial 3: 28.54−3.40=25.14 mL28.54 - 3.40 = 25.14\text{ mL}. Trial 3 differs from the average of Trials 1 and 2 by only +0.04 mL+0.04\text{ mL}, which is within normal burette tolerance (pm0.05 mL\\pm 0.05\text{ mL}), so Trial 3 should be retained (or the bubble was negligible).

    While an air bubble dislodging during titration does theoretically overestimate delivered titrant, you must inspect the actual delivered volume (25.14 mL25.14\text{ mL}) compared to Trials 1 and 2 (25.12 mL25.12\text{ mL} and 25.08 mL25.08\text{ mL}). Because the deviation (+0.04 mL+0.04\text{ mL}) falls within standard random reading error, the bubble did not cause a significant systematic error.

  2. Step 2

    n(KHP)=0.5106 g/204.22 g/mol=0.002500 mol  ⟹  Moles of NaOH=0.002500 moln(\text{KHP}) = 0.5106\text{ g} / 204.22\text{ g/mol} = 0.002500\text{ mol} \implies \text{Moles of NaOH} = 0.002500\text{ mol}

  3. Step 3

    V(solution)=50.0 mL=0.0500 L  ⟹  M=0.002500 mol/0.0500 L=0.0500 MV(\text{solution}) = 50.0\text{ mL} = 0.0500\text{ L} \implies M = 0.002500\text{ mol} / 0.0500\text{ L} = 0.0500\text{ M}

    Correct

    C=nNaOHVNaOH,delivered=0.0025002 mol0.02510 L≈0.0996 MC = \frac{n_{\ce{NaOH}}}{V_{\ce{NaOH},\text{delivered}}} = \frac{0.0025002\text{ mol}}{0.02510\text{ L}} \approx 0.0996\text{ M}

    You divided the moles of NaOH\ce{NaOH} by the volume of deionized water (50.0 mL50.0\text{ mL}) used to dissolve the KHP in the Erlenmeyer flask. To find the concentration of the NaOH\ce{NaOH} titrant, you must divide by the delivered volume of NaOH\ce{NaOH} measured from the burette readings (approximately 25.10 mL=0.02510 L25.10\text{ mL} = 0.02510\text{ L}).

  4. Step 4

    The nominal range is 0.0990 M0.0990\text{ M} to 0.1010 M0.1010\text{ M}. 0.0500 M0.0500\text{ M} is outside the range, so the experimental result does not agree.

    Correct

    The calculated mean concentration (0.09956 M0.09956\text{ M} to 0.09961 M0.09961\text{ M}) falls within [0.0990 M,0.1010 M][0.0990\text{ M}, 0.1010\text{ M}], confirming agreement with the manufacturer's claim.

    Because your calculated concentration of 0.0500 M0.0500\text{ M} was based on the dissolution volume rather than the delivered titrant volume, your final comparison with the manufacturer specification was affected.

WHAT YOU DID WELL
Compliments from Your Teacher

You correctly calculated the moles of KHP using its molar mass.

You correctly applied the 1:1 reaction stoichiometry to relate moles of KHP to moles of NaOH.

You clearly stated the manufacturer's tolerance range based on the given nominal claim.

THE ERROR
Where Your ACEpath Teacher Spotted the Mistake
Incorrect Volume in Titrant Molarity Calculation

no

WHY IT HAPPENED
ACEpath Teacher's Diagnosis

Confusing the volume of solvent used to dissolve the primary standard in the receiving flask with the volume of titrant dispensed from the burette.

REASONING PATH
Your ACEpath teacher’s suggested approach
Choosing the Correct Volume for Molarity
  • Molarity of titrant is defined as moles of titrant÷volume of titrant delivered from the burette\text{moles of titrant} \div \text{volume of titrant delivered from the burette}.

  • Never use the volume of deionized water in the receiving flask to calculate titrant molarity; water volume affects neither the moles of primary standard nor the titrant concentration.

Verifying Systematic Errors with Data
  • A suspected procedural flaw (like an air bubble) must be corroborated by comparing the delivered titer to other concordant trials.

  • Always compute the delivered burette volume (Vf−ViV_f - V_i) for every trial before deciding whether to reject a suspected outlier.

CORRECTED PATH
ACEpath Teacher's Demonstration
  1. Delivered burette volumes: Trial 1 = 25.62−0.50=25.12 mL25.62 - 0.50 = 25.12\text{ mL}; Trial 2 = 26.23−1.15=25.08 mL26.23 - 1.15 = 25.08\text{ mL}; Trial 3 = 28.54−3.40=25.14 mL28.54 - 3.40 = 25.14\text{ mL}.

  2. Mean delivered volume of valid trials: V‾=25.12+25.08+25.143=25.113 mL=0.025113 L\overline{V} = \frac{25.12 + 25.08 + 25.14}{3} = 25.113\text{ mL} = 0.025113\text{ L} (or for Trials 1 and 2: 25.10 mL=0.02510 L25.10\text{ mL} = 0.02510\text{ L}).

  3. Mean molarity: M=0.0025002 mol0.025113 L=0.09956 MM = \frac{0.0025002\text{ mol}}{0.025113\text{ L}} = 0.09956\text{ M} (or 0.09961 M0.09961\text{ M} using Trials 1 and 2).

  4. Comparison: 0.09956 M0.09956\text{ M} lies within [0.0990 M,0.1010 M][0.0990\text{ M}, 0.1010\text{ M}], so the result agrees with the manufacturer's specification.

Final answer

Trial 3 may be retained because its delivered volume (25.14 mL25.14\text{ mL}) is within normal reading variation (+0.04 mL+0.04\text{ mL} of the Trial 1–2 average). The mean concentration of NaOH\ce{NaOH} is 0.0996 M0.0996\text{ M} (or 0.09956 M0.09956\text{ M} across all trials), which agrees with the manufacturer's claim of 0.1000 M±0.0010 M0.1000\text{ M} \pm 0.0010\text{ M}.

NEXT STEPS
Final Takeaways from Your Teacher
  • In future titration problems, draw a clear distinction between analyte flask contents (moles of solute) and burette contents (titrant volume delivered).

  • Always calculate the delivered volumes (Vf−ViV_f - V_i) for all trials first to evaluate concordance numerically before discarding any run.

OverviewACEpath Teacher's Diagnosis

03 The student uses only the final burette readings without subtracting the initial volumes to determine the delivered titrant volume.

Sample Student Work

Reviewed by ACEpath Teacher
Step-by-Step Review
Walk Through It with Your Teacher
Incorrect
  1. Step 1

    Trial 3 is discarded because an air bubble escaping during delivery causes an artificially high volume reading.

    Correct

    V1=25.12 mLV_1 = 25.12\text{ mL}, V2=25.08 mLV_2 = 25.08\text{ mL}, V3=25.14 mLV_3 = 25.14\text{ mL}; because V3V_3 is within 0.04 mL0.04\text{ mL} of the mean of Trials 1 and 2, the bubble caused no significant systematic error, so Trial 3 can be retained (or excluded conservatively without altering the conclusion).

    While an escaping air bubble theoretically inflates the recorded delivery volume, you must verify the actual net delivered volume from the data. Here, ΔV3=28.54 mL−3.40 mL=25.14 mL\Delta V_3 = 28.54\text{ mL} - 3.40\text{ mL} = 25.14\text{ mL}, which is fully concordant with Trials 1 and 2 (25.12 mL25.12\text{ mL} and 25.08 mL25.08\text{ mL}).

  2. Step 2

    n(KHP)=0.5106 g/204.22 g/mol=0.002500 moln(\text{KHP}) = 0.5106\text{ g} / 204.22\text{ g/mol} = 0.002500\text{ mol}

  3. Step 3

    V1=25.62 mL=0.02562 L  ⟹  V2=26.23 mL=0.02623 LV_1 = 25.62\text{ mL} = 0.02562\text{ L} \implies V_2 = 26.23\text{ mL} = 0.02623\text{ L}

    Correct

    V1=25.62 mL−0.50 mL=25.12 mL=0.02512 L  ⟹  V2=26.23 mL−1.15 mL=25.08 mL=0.02508 LV_1 = 25.62\text{ mL} - 0.50\text{ mL} = 25.12\text{ mL} = 0.02512\text{ L} \implies V_2 = 26.23\text{ mL} - 1.15\text{ mL} = 25.08\text{ mL} = 0.02508\text{ L}

    You took the final burette reading directly as the delivered volume instead of subtracting the initial reading (Vdelivered=Vfinal−VinitialV_{\text{delivered}} = V_{\text{final}} - V_{\text{initial}}).

  4. Step 4

    M1=0.002500 mol/0.02562 L=0.09758 M  ⟹  M2=0.002500 mol/0.02623 L=0.09531 M  ⟹  Mean M=(0.09758+0.09531)/2=0.09645 MM_1 = 0.002500\text{ mol} / 0.02562\text{ L} = 0.09758\text{ M} \implies M_2 = 0.002500\text{ mol} / 0.02623\text{ L} = 0.09531\text{ M} \implies \text{Mean } M = (0.09758 + 0.09531) / 2 = 0.09645\text{ M}

    Correct

    M1=2.5002×10−3 mol0.02512 L=0.09953 M  ⟹  M2=2.5002×10−3 mol0.02508 L=0.09969 M  ⟹  Mean M=0.09961 MM_1 = \frac{2.5002 \times 10^{-3}\text{ mol}}{0.02512\text{ L}} = 0.09953\text{ M} \implies M_2 = \frac{2.5002 \times 10^{-3}\text{ mol}}{0.02508\text{ L}} = 0.09969\text{ M} \implies \text{Mean } M = 0.09961\text{ M} (or 0.09956 M0.09956\text{ M} across all three trials)

    Your molarity calculations and averaging arithmetic followed the correct formulas, but the numerical values were skewed by using uncorrected burette readings.

  5. Step 5

    The nominal range is 0.0990 M0.0990\text{ M} to 0.1010 M0.1010\text{ M}. Since 0.09645 M0.09645\text{ M} is below 0.0990 M0.0990\text{ M}, the experimental result does not agree.

    Correct

    The calculated mean (0.09961 M0.09961\text{ M}, or 0.09956 M0.09956\text{ M}) falls within [0.0990 M,0.1010 M][0.0990\text{ M}, 0.1010\text{ M}], so the experimental result agrees with the manufacturer's claim.

    Because your calculated mean concentration was artificially low due to the volume error, you reached the opposite conclusion regarding agreement with the manufacturer's specification.

WHAT YOU DID WELL
Compliments from Your Teacher

You correctly calculated the moles of primary standard KHP using its molar mass.

You correctly identified that a 1:11:1 stoichiometric ratio relates KHP and NaOH.

You correctly set up the formula for molarity as moles divided by volume in liters.

You correctly identified the acceptable tolerance interval for the manufacturer's specification.

THE ERROR
Where Your ACEpath Teacher Spotted the Mistake
Omission of Initial Burette Reading

no

WHY IT HAPPENED
ACEpath Teacher's Diagnosis

You treated the final burette scale reading as the delivered titrant volume rather than calculating the difference between the final and initial readings (ΔV=Vfinal−Vinitial\Delta V = V_{\text{final}} - V_{\text{initial}}).

REASONING PATH
Your ACEpath teacher’s suggested approach
Net Volume Delivery
  • A burette measures volume by difference: Vdelivered=Vfinal−VinitialV_{\text{delivered}} = V_{\text{final}} - V_{\text{initial}}.

  • Always record both initial and final levels, and subtract VinitialV_{\text{initial}} from VfinalV_{\text{final}} before using the volume in stoichiometric calculations.

Evaluating Systematic Errors with Data
  • Theoretical procedural issues (such as tip bubbles) must be evaluated against the quantitative data of all trials.

  • Calculate the net delivered volume ΔV\Delta V for every trial first; discard a trial only if its net volume significantly deviates from concordant replicates.

CORRECTED PATH
ACEpath Teacher's Demonstration
  1. Calculate delivered volumes: V1=25.62 mL−0.50 mL=25.12 mL=0.02512 LV_1 = 25.62\text{ mL} - 0.50\text{ mL} = 25.12\text{ mL} = 0.02512\text{ L}, V2=26.23 mL−1.15 mL=25.08 mL=0.02508 LV_2 = 26.23\text{ mL} - 1.15\text{ mL} = 25.08\text{ mL} = 0.02508\text{ L}, and V3=28.54 mL−3.40 mL=25.14 mL=0.02514 LV_3 = 28.54\text{ mL} - 3.40\text{ mL} = 25.14\text{ mL} = 0.02514\text{ L}.

  2. Calculate molarity for each trial using M=2.5002×10−3 molVM = \frac{2.5002 \times 10^{-3}\text{ mol}}{V}: M1=0.09953 MM_1 = 0.09953\text{ M}, M2=0.09969 MM_2 = 0.09969\text{ M}, and M3=0.09945 MM_3 = 0.09945\text{ M}.

  3. Calculate the average concentration: M‾1−2=0.09961 M\overline{M}_{1-2} = 0.09961\text{ M} (or M‾all=0.09956 M\overline{M}_{\text{all}} = 0.09956\text{ M}).

  4. Compare to the manufacturer's range: both values lie within [0.0990 M,0.1010 M][0.0990\text{ M}, 0.1010\text{ M}], confirming agreement.

Final answer

The mean concentration is 0.09961 M0.09961\text{ M} (using Trials 1 and 2) or 0.09956 M0.09956\text{ M} (using all three trials); both agree with the manufacturer's claim of 0.1000 M±0.0010 M0.1000\text{ M} \pm 0.0010\text{ M}.

NEXT STEPS
Final Takeaways from Your Teacher
  • Always subtract the initial burette reading from the final burette reading to find delivered volume.

  • Compute net volumes for all titration trials before deciding whether to retain or discard a suspect run.

  • Double-check that calculated concentrations fall within expected physical ranges relative to nominal values.

Lock it in

Similar question

01

A student rinses a burette with deionized water but forgets to rinse it with the standard HCl\text{HCl} titrant before filling. Explain whether this is a systematic or random error, predict the effect on the recorded titer volume, and state whether repeating the titration five times under the same conditions will eliminate the error.

Answer

This is a systematic error. Residual water dilutes the HCl\text{HCl} titrant, requiring a larger volume of titrant to reach the endpoint (artificially high titer). Because the bias is unidirectional, averaging repeated trials under the same flawed conditions will not eliminate the error.

02

In a redox titration, the concordant volumes for Trials 1 and 2 are 21.40 mL21.40\text{ mL} and 21.45 mL21.45\text{ mL}. In Trial 3, a small splash of analyte left the flask during swirling, and the recorded volume was 20.80 mL20.80\text{ mL}. Should Trial 3 be included in calculating the mean concentration? Justify your decision using error analysis.

Answer

Trial 3 must be discarded. Splashing analyte out of the flask is a known systematic error (determinate loss of moles), which directly caused the significantly lower titer (20.80 mL20.80\text{ mL} vs. mean 21.425 mL21.425\text{ mL}, deviation −0.625 mL-0.625\text{ mL}). Averaging an acknowledged, unrepresentative run biases the result.

03

A student standardizes NaOH\text{NaOH} with primary-standard KHP (M=204.22 g/molM = 204.22\text{ g/mol}). A 0.4084 g0.4084\text{ g} sample requires 20.10 mL20.10\text{ mL} of NaOH\text{NaOH}. If the analytical balance used had a persistent tare offset that read +0.0100 g+0.0100\text{ g} higher than the true mass, calculate the true mass of KHP, the true concentration of NaOH\text{NaOH}, and the percentage error introduced into the molarity by this systematic weighing error.

Answer
  1. True mass =0.4084 g−0.0100 g=0.3984 g= 0.4084\text{ g} - 0.0100\text{ g} = 0.3984\text{ g}.\n2. True moles =0.3984204.22=1.9508×10−3 mol= \frac{0.3984}{204.22} = 1.9508 \times 10^{-3}\text{ mol}.\n3. True [extNaOH]=1.9508×10−3 mol0.02010 L=0.09706 M[ ext{NaOH}] = \frac{1.9508 \times 10^{-3}\text{ mol}}{0.02010\text{ L}} = 0.09706\text{ M}.\n4. Apparent [extNaOH]=0.4084/204.220.02010=0.09949 M[ ext{NaOH}] = \frac{0.4084 / 204.22}{0.02010} = 0.09949\text{ M}.\n5. Percentage error =0.09949−0.097060.09706×100%=+2.50%= \frac{0.09949 - 0.09706}{0.09706} \times 100\% = +2.50\%.